Performance of a cylindrical diode array for use in a 1.5 T MR-linac
- 15 January 2016
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 61 (3) , N80-N89
Abstract
At the UMC Utrecht, a linear accelerator with integrated magnetic resonance imaging (MRI) has been developed, the MR-linac. Patient-specific quality assurance (QA) of treatment plans for MRI-based image guided radiotherapy requires QA equipment compatible with this 1.5 T magnetic field. The purpose of this study was to examine the performance characteristics of the ArcCHECK-MR in a transverse 1.5 T magnetic field. To this end, the short-term reproducibility, dose linearity, dose rate dependence, field size dependence, dose per pulse dependence and inter-diode dose response variation of the ArcCHECK-MR diode array were evaluated on a conventional linac and on the MR-linac. The ArcCHECK-MR diode array performed well for all tests on both linacs, no significant differences in performance characteristics were observed. Differences in the maximum dose deviations between both linacs were less than 1.5%. Therefore, we conclude that the ArcCHECK-MR can be used in a transverse 1.5 T magnetic field.Keywords
This publication has 11 references indexed in Scilit:
- Patient-Specific Quality Assurance for the Delivery of 60Co Intensity Modulated Radiation Therapy Subject to a 0.35-T Lateral Magnetic FieldInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Relative dosimetry in a 1.5 T magnetic field: an MR-linac compatible prototype scanning water phantomPhysics in Medicine & Biology, 2014
- The Australian Magnetic Resonance Imaging–Linac ProgramSeminars in Radiation Oncology, 2014
- The Magnetic Resonance Imaging–Linac SystemSeminars in Radiation Oncology, 2014
- Performance of a multi-axis ionization chamber array in a 1.5 T magnetic fieldPhysics in Medicine & Biology, 2014
- A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systemsRadiotherapy and Oncology, 2013
- Evaluation of the ArcCHECK QA system for IMRT and VMAT verificationPhysica Medica, 2013
- Integrated megavoltage portal imaging with a 1.5 T MRI linacPhysics in Medicine & Biology, 2011
- Integrating a 1.5 T MRI scanner with a 6 MV accelerator: proof of conceptPhysics in Medicine & Biology, 2009
- First MR images obtained during megavoltage photon irradiation from a prototype integrated linac‐MR systemMedical Physics, 2009