Comparison of Repeatability and Agreement between Swept-Source Optical Biometry and Dual-Scheimpflug Topography
Open Access
- 10 December 2017
- journal article
- research article
- Published by Wiley in Journal of Ophthalmology
- Vol. 2017, 1-5
Abstract
Purpose. To assess the repeatability and agreement of parameters obtained with two biometers and to compare the predictability. Methods. Biometry was performed on 101 eyes with cataract using the IOLMaster 700 and the Galilei G6. Three measurements were obtained per eye with each device, and repeatability was evaluated. The axial length (AL), anterior chamber depth (ACD), keratometry (K), white-to-white (WTW) corneal diameter, central corneal thickness (CCT), and lens thickness (LT) were measured and postoperative predictability was compared. Results. Measurements could not be obtained with the IOLMaster 700 in one eye and in seven eyes with the Galilei G6 due to dense cataract. Both the IOLMaster 700 and Galilei G6 showed good repeatability, although the IOLMaster 700 showed better repeatability than the Galilei G6. There were no statistically significant differences in AL, ACD, steepest K, WTW, and LT (), although flattest K, mean K, and CCT differed (). The proportion of eyes with an absolute prediction error within 0.5 D was 85.0% for the IOLMaster 700 and was 80.0% for the Galilei G6 based on the SRK/T formula. Conclusions. Two biometers showed high repeatability and relatively good agreements. The swept-source optical biometer demonstrated better repeatability, penetration, and an overall lower prediction error.
Keywords
Funding Information
- Inha University
This publication has 16 references indexed in Scilit:
- Biometry with a new swept-source optical coherence tomography biometer: Repeatability and agreement with an optical low-coherence reflectometry deviceJournal of Cataract & Refractive Surgery, 2016
- Comparison of Two Optical BiometersOptometry and Vision Science, 2016
- Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometryJournal of Cataract & Refractive Surgery, 2016
- Clinical comparison of a new swept-source optical coherence tomography–based optical biometer and a time-domain optical coherence tomography–based optical biometerJournal of Cataract & Refractive Surgery, 2015
- Reproducibility of a Long-Range Swept-Source Optical Coherence Tomography Ocular Biometry System and Comparison with Clinical BiometersOphthalmology, 2013
- Comparison of Corneal Tomography Measurements Using Galilei, Orbscan II, and Placido Disk–based Topographer SystemsJournal of Refractive Surgery, 2011
- Comparison of central corneal thickness measurements with the Galilei dual Scheimpflug analyzer and ultrasound pachymetryIndian Journal Of Ophthalmology, 2010
- Fourier-domain biophotoacoustic subsurface depth selective amplitude and phase imaging of turbid phantoms and biological tissueJournal of Biomedical Optics, 2006
- Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliabilityJournal of Cataract & Refractive Surgery, 2001
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992