Noise reduction in MR angiography with nonlinear anisotropic filtering
Open Access
- 21 April 2004
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 19 (5) , 632-639
Abstract
Purpose To evaluate three-dimensional nonlinear anisotropic filtering in suppressing image noise in high spatial resolution magnetic resonance angiograms (MRA) acquired with hybrid undersampled projection reconstruction and phase contrast vastly undersampled isotropic projection reconstruction (PC-VIPR). Materials and Methods Three-dimensional nonlinear anisotropic filtering was quantitatively analyzed and evaluated through the measurement of contrast to noise ratio (CNR) in PC-VIPR images and contrast enhanced peripheral MRA images. To filter MRA images with ultra-high spatial resolution and poor CNR, a spatial frequency dependent nonlinear anisotropic filtering algorithm was proposed that uses two-step processing to filter the whole spatial frequency data. Results Three-dimensional nonlinear anisotropic filtering was shown to be effective in suppressing noise and improving CNR in MRA with isotropic spatial resolution. Higher CNR was achieved using spatial frequency dependent nonlinear anisotropic filtering. A typical CNR gain of between 50–100% was shown in our studies. Conclusion Three-dimensional nonlinear anisotropic filtering significantly improved CNR in MRA images with isotropic spatial resolution. Spatial frequency dependent nonlinear anisotropic filtering further improved CNR for MRA images with ultra-high spatial resolution and low CNR. J. Magn. Reson. Imaging 2004;19:632–639.Keywords
This publication has 17 references indexed in Scilit:
- Time‐resolved, undersampled projection reconstruction imaging for high‐resolution CE‐MRA of the distal runoff vesselsMagnetic Resonance in Medicine, 2002
- Time‐resolved contrast‐enhanced imaging with isotropic resolution and broad coverage using an undersampled 3D projection trajectoryMagnetic Resonance in Medicine, 2002
- 3D contrast‐enhanced MR angiography of the run‐off vessels: Value of image subtractionJournal of Magnetic Resonance Imaging, 2001
- Vessel enhancement filtering in three‐dimensional MR angiograms using long‐range signal correlationJournal of Magnetic Resonance Imaging, 1997
- The study of logarithmic image processing model and its application to image enhancementIEEE Transactions on Image Processing, 1995
- A multidimensional nonlinear edge-preserving filter for magnetic resonance image restorationIEEE Transactions on Image Processing, 1995
- Reduction of partial‐volume artifacts with zero‐filled interpolation in three‐dimensional MR angiographyJournal of Magnetic Resonance Imaging, 1994
- Nonlinear anisotropic filtering of MRI dataIEEE Transactions on Medical Imaging, 1992
- Selection of a convolution function for Fourier inversion using gridding (computerised tomography application)IEEE Transactions on Medical Imaging, 1991
- Biased anisotropic diffusion: a unified regularization and diffusion approach to edge detectionImage and Vision Computing, 1990