High-dynamic-range micro-CT for nondestructive testing of titanium 3D-printed medical components
Open Access
- 1 July 2022
- journal article
- research article
- Published by SPIE-Intl Soc Optical Eng in Journal of Medical Imaging
- Vol. 9 (04) , 044004-13
Abstract
Purpose: Industrial microcomputed tomography (micro-CT) scanners are suitable for nondestructive testing (NDT) of metal, 3D-printed medical components. Typically, these scanners are equipped with high-energy sources that require heavy shielding and costly infrastructure to operate safely, making routine NDT of medical components prohibitively expensive. Alternatively, fixed-current, low-cost x-ray units could be implemented to perform CT-based NDT of 3D-printed medical parts in a subset of cases, if there is sufficient x-ray transmission for the CT reconstruction. A lack of signal—caused by areas of high attenuation in two-dimensional-projection images of metal objects—leads to artifacts that can make an image-based NDT unreliable. We present the implementation of a dual-exposure technique devised to extend the dynamic range (DR) of a commercially available CT scanner equipped with a low-cost low-energy (80 kV) x-ray unit, increasing the signal-to-noise ratio of highly attenuated areas for NDT of 3D-printed medical components.Approach: Our high-dynamic-range CT (HDR-CT) technique adequately combines projection images acquired at two exposure levels by modifying the integration times of each protocol. We evaluate the performance and limitations of this HDR-CT technique by imaging a series of titanium-alloy test-samples. One of the test-samples was a resolution and conspicuity phantom designed to assess the improvements in void visualization of the proposed methodology. The other test-samples were four porous cylinders, 17 × 40 mm, with 60%, 70%, 80%, and 90% nominal internal porosities.Results: Our HDR-CT technique adequately combines projection images acquired at two exposure levels by modifying the integration times of each protocol. Our results demonstrate that the 12-bit native DR of the CT scanner was increased to effective values of between 14 and 16 bits.Conclusions: The HDR-CT reconstructions showed improved contrast-to-noise and void conspicuity, when compared with conventional CT scans. This extension of DR has the potential to improve defect visualization during NDT of medium-size, titanium-alloy, 3D-printed medical components.Keywords
This publication has 17 references indexed in Scilit:
- Enhanced dynamic range x-ray imagingComputers in Biology and Medicine, 2017
- Comparison of medical and industrial X-ray computed tomography for non-destructive testingCase Studies in Nondestructive Testing and Evaluation, 2016
- Varying-energy CT imaging method based on EM-TVMeasurement Science and Technology, 2016
- Improved contrast of materials based on multi-voltage images decomposition in X-ray CTMeasurement Science and Technology, 2015
- High-Dynamic-Range CT Reconstruction Based on Varying Tube-Voltage ImagingPLOS ONE, 2015
- Industrial applications of computed tomographyCIRP Annals, 2014
- Dual-exposure technique for extending the dynamic range of x-ray flat panel detectorsPhysics in Medicine & Biology, 2013
- Stripe and ring artifact removal with combined wavelet—Fourier filteringOptics Express, 2009
- A method for extending the dynamic range of flat panel imagers for use in cone beam computed tomography2005
- Optimal short scan convolution reconstruction for fan beam CTMedical Physics, 1982