Upgrade of X-ray crystal spectrometer for high temperature measurement using neon-like xenon lines on EAST
Authors:
- 21 September 2018
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 89 (10) , 10F110
Abstract
A two-crystal X-ray spectrometer system has been implemented in the EAST tokamak to simultaneously diagnose high- and low-temperature plasmas using He- and H-like argon spectra. But for future fusion devices like ITER and Chinese Fusion Engineering Test Reactor (CFETR), argon ions become fully stripped in the core and the intensity of the H-like lines will be significantly at high temperatures (Te > 5 keV). With increasing auxiliary heating power on EAST, the core plasma temperature could also reach 5 keV and higher. In such conditions, the use of a xenon puff becomes an appropriate choice for both ion-temperature and flow-velocity measurements. A new two-crystal system using a quartz 110 crystal (2d = 4.913 Å) to view He-like argon lines and a quartz 011 crystal (2d = 6.686 Å) to view Ne-like xenon spectra has been deployed on a poloidal X-ray crystal spectrometer. While the He-like argon spectra will be used to measure the plasma temperature in the edge plasma region, the Ne-like xenon spectra will be used for measurement in the hot core. The new crystal arrangement allows a wide temperature measurement ranging from 0.5 to 10 keV or even higher, being the first tests for burning plasmas like ITER and CFETR. The preliminary result of lab-tests, Ne-like xenon lines measurement will be presented.Keywords
This publication has 12 references indexed in Scilit:
- Overview of EAST experiments on the development of high-performance steady-state scenarioNuclear Fusion, 2017
- Upgrades of poloidal and tangential x-ray imaging crystal spectrometers for temperature and rotation measurements on EASTReview of Scientific Instruments, 2016
- Measurement of helium-like and hydrogen-like argon spectra using double-crystal X-ray spectrometers on EASTReview of Scientific Instruments, 2016
- Upgrades of imaging x-ray crystal spectrometers for high-resolution and high-temperature plasma diagnostics on EASTReview of Scientific Instruments, 2014
- Validation of x-ray line ratios for electron temperature determination in tokamak plasmasJournal of Physics B: Atomic, Molecular and Optical Physics, 2014
- The first experimental results from x-ray imaging crystal spectrometer for KSTARReview of Scientific Instruments, 2010
- Objectives and layout of a high-resolution x-ray imaging crystal spectrometer for the large helical deviceReview of Scientific Instruments, 2010
- Imaging x-ray crystal spectrometer on EASTPlasma Physics and Controlled Fusion, 2010
- Spatially resolved high resolution x-ray spectroscopy for magnetically confined fusion plasmas (invited)Review of Scientific Instruments, 2008
- Results from the National Spherical Torus Experiment x-ray crystal spectrometerReview of Scientific Instruments, 2003