Reproducibility of DIAGNOdent 2095 and DIAGNOdent Pen measurements: results from an in vitro study on occlusal sites
Authors:
- Jan KühnischJan Kühnisch
- Katharina BücherKBKatharina Bücher
- Volkmar HenschelVHVolkmar Henschel
- Reinhard HickelRHReinhard Hickel
- 20 June 2007
- journal article
- research article
- Published by Wiley in European Journal of Oral Sciences
- Vol. 115 (3) , 206-211
Abstract
The aim of this in vitro study was to investigate the intra‐ and interdevice reproducibility of the DIAGNOdent 2095 (DD2095) and the new DIAGNOdent Pen (DDPen) on non‐cavitated occlusal caries sites. Altogether, 241 sites on 90 molars were inspected twice with both devices and by four examiners. The intra/interdevice reproducibilities were assessed by means of the intraclass correlation coefficient (ICC) as well as by the range of the 95% limits of agreement of Bland & Altman. Additionally, a linear mixed‐effects model was fitted with fixed effects for the visual scores and the devices used and random effects for occlusal sites and examiners within each measurement site. In result, the following values were determined for the intradevice reproducibility: DD2095: ICC 0.89, range 42.3; and DDPen: ICC 0.88, range 49.3. The interdevice reproducibility was found to be in the same order of magnitude (ICC 0.82, range 53.7). The comparison between both devices indicated significantly lower mean values for the DDPen. The estimates of the linear model show that there is a high variation between the sites that cannot be explained by the fixed effects. In conclusion, both devices showed an imperfect reproducibility, which indicate the usage as adjunct tool only in clinical practice.Keywords
This publication has 24 references indexed in Scilit:
- Influence of Different Storage Methods on Laser Fluorescence Values: A Two-Year StudyCaries Research, 2006
- An in vitro study of the reliability of DIAGNOdent® measurementsJournal of Oral Rehabilitation, 2004
- Consistency of DIAGNOdent instruments for clinical assessment of fissure cariesActa Odontologica Scandinavica, 2004
- In vivo Effectiveness of Laser Fluorescence Compared to Visual Inspection and Radiography for the Detection of Occlusal Caries in Primary TeethCaries Research, 2003
- Performance of Conventional and New Methods for the Detection of Occlusal Caries in Deciduous TeethCaries Research, 2003
- Clinical Study of the Use of the Laser Fluorescence Device DIAGNOdent for Detection of Occlusal Caries in ChildrenCaries Research, 2003
- Validity and Reproducibility of a Laser Fluorescence System for Detecting the Activity of White-Spot Lesions on Free Smooth Surfaces in vivoCaries Research, 2002
- Occlusal Caries Detection with KaVo DIAGNOdent and Radiography: An in vitro ComparisonCaries Research, 2000
- Performance and Reproducibility of a Laser Fluorescence System for Detection of Occlusal Caries in vitroCaries Research, 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986