An analytical form of ring artifact correction for computed tomography based on directional gradient domain optimization
Authors:
- 7 March 2024
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 51 (6) , 4121-4132
Abstract
Background: Ring artifact is a common problem in Computed Tomography (CT), which can lead to inaccurate diagnoses and treatment plans. It can be caused by various factors such as detector imperfections, anti‐scatter grids, or other nonuniform filters placed in the x‐ray beam. Physics‐based corrections for these x‐ray source and detector non‐uniformity, in general cannot completely get rid of the ring artifacts. Therefore, there is a need for a robust method that can effectively remove ring artifacts in the image domain while preserving details.Purpose: This study aims to develop an effective method for removing ring artifacts from reconstructed CT images.Methods: The proposed method starts by converting the reconstructed CT image containing ring artifacts into polar coordinates, thereby transforming these artifacts into stripes. Relative Total Variation is used to extract the image's overall structural information. For the efficient restoration of intricate details, we introduce Directional Gradient Domain Optimization (DGDO) and design objective functions that make use of both the image's gradient and its overall structure. Subsequently, we present an efficient analytical algorithm to minimize these objective functions. The image obtained through DGDO is then transformed back into Cartesian coordinates, finalizing the ring artifact correction process.Results: Through a series of synthetic and real‐world experiments, we have effectively demonstrated the prowess of our proposed method in the correction of ring artifacts while preserving intricate details in reconstructed CT images. In a direct comparison, our method has exhibited superior visual quality compared to several previous approaches. These results underscore the remarkable potential of our approach for enhancing the overall quality and clinical utility of CT imaging.Conclusions: The proposed method offers an analytical solution for removing ring artifacts from CT images while preserving details. As ring artifacts are a common problem in CT imaging, this method has high practical value in the medical field. The proposed method can improve image quality and reduce the difficulty of disease diagnosis, thereby contributing to better patient care.Keywords
Funding Information
- National Natural Science Foundation of China (62031020)
This publication has 33 references indexed in Scilit:
- A new iterative algorithm for ring artifact reduction in CT using ring total variationMedical Physics, 2019
- Removing ring artifacts in cone‐beam CT via TV‐Stokes and unidirectional total variation modelMedical Physics, 2019
- Removing ring artefacts in CT images via unidirectional relative variation modelElectronics Letters, 2016
- High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imagingPhysics in Medicine & Biology, 2011
- Classification of ring artifacts for their effective removal using type adaptive correction schemesComputers in Biology and Medicine, 2011
- Removal of ring artifacts in CT imaging through detection and correction of stripes in the sinogramPhysics in Medicine & Biology, 2010
- Ring artifact correction for high-resolution micro CTPhysics in Medicine & Biology, 2009
- Reduction of ring artefacts in high resolution micro-CT reconstructionsPhysics in Medicine & Biology, 2004
- Numerical removal of ring artifacts in microtomographyReview of Scientific Instruments, 1998
- X-ray microtomography scanner using time-delay integration for elimination of ring artefacts in the reconstructed imageNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997