A method for estimating radiation interaction coefficients for tissues from single energy CT
Author:
- 13 November 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (23) , 7479-7499
Abstract
A parametric model for the x-ray linear attenuation coefficient is used to describe the compositional dependence of Hounsfield numbers measured by medical CT scanners. Measurements with materials of known density and composition, that span and evenly sample the compositional range of tissues, are written as linear simultaneous equations and solved for model coefficients. An algorithm is identified for this purpose. Results are expressed as atomic cross-sections in units of barn per electron divided by the attenuation coefficient for water. With the CT scanner characterised, a virtual CT scan can be simulated to predict HN for tissues based upon their known density and composition. Similar calculations using the tabulations and mixture rule deliver attenuation coefficients and mass energy absorption coefficients for mono-energetic radiation 10 keV to 20 MeV. Results are presented for measurements with a radiotherapy CT simulator, the RMI-467 phantom with tissue substitute materials, plus common polymer materials and silicon. Published measurements with earlier generations of the phantom and tissue substitutes using different CT scanners are also considered. Measured atomic cross-sections differ from expectations for mono-energetic radiation due to the use of a filtered spectrum and energy integrating detection system. The cross-sections for different CT scanners are similar, without large variations with kVp. Results are presented showing the relationship between predicted HN for tissues, electron density and photon interaction coefficients for healthy tissues and mono-energetic radiation. A strategy is suggested for accommodating strongly attenuating materials such as calculi and metallic implants.Keywords
This publication has 28 references indexed in Scilit:
- Calibration of CT Hounsfield units for proton therapy treatment planning: use of kilovoltage and megavoltage images and comparison of parameterized methodsPhysics in Medicine & Biology, 2013
- Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculationsPhysics in Medicine & Biology, 2008
- Voxel-based computational models of real human anatomy: a reviewRadiation and Environmental Biophysics, 2004
- A parameterization scheme for the x-ray linear attenuation coefficient and energy absorption coefficientPhysics in Medicine & Biology, 2004
- Attenuation correction for a combined 3D PET/CT scannerMedical Physics, 1998
- Comparison of x‐ray cross sections for diagnostic and therapeutic medical physicsMedical Physics, 1996
- An electron density calibration phantom for CT-based treatment planning computersMedical Physics, 1992
- An algorithm for linear least squares problems with equality and nonnegativity constraintsMathematical Programming, 1981
- X-ray attenuation coefficients of elements and mixturesPhysics Reports, 1981
- Energy-selective reconstructions in X-ray computerised tomographyPhysics in Medicine & Biology, 1976