Design and application of a four‐channel transmit/receive surface coil for functional cardiac imaging at 7T
Authors:
- Matthias A. DieringerMDMatthias A. Dieringer
- Tomasz LindelTLTomasz Lindel
- Helmar WaicziesHWHelmar Waiczies
- Jan RiegerJRJan Rieger
- Harald PfeifferHPHarald Pfeiffer
Open Access
- 1 February 2011
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 33 (3) , 736-741
Abstract
Purpose: To design and evaluate a four‐channel cardiac transceiver coil array for functional cardiac imaging at 7T.Materials and Methods: A four‐element cardiac transceiver surface coil array was developed with two rectangular loops mounted on an anterior former and two rectangular loops on a posterior former. specific absorption rate (SAR) simulations were performed and a B calibration method was applied prior to obtain 2D FLASH CINE (mSENSE, R = 2) images from nine healthy volunteers with a spatial resolution of up to 1 × 1 × 2.5 mm3.Results: Tuning and matching was found to be better than 10 dB for all subjects. The decoupling (S21) was measured to be >18 dB between neighboring loops, >20 dB for opposite loops, and >30 dB for other loop combinations. SAR values were well within the limits provided by the IEC. Imaging provided clinically acceptable signal homogeneity with an excellent blood‐myocardium contrast applying the B calibration approach.Conclusion: A four‐channel cardiac transceiver coil array for 7T was built, allowing for cardiac imaging with clinically acceptable signal homogeneity and an excellent blood‐myocardium contrast. Minor anatomic structures, such as pericardium, mitral, and tricuspid valves and their apparatus, as well as trabeculae, were accurately delineated. J. Magn. Reson. Imaging 2011;.Keywords
This publication has 17 references indexed in Scilit:
- Quantitative assessment of left ventricular function in humans at 7 TMagnetic Resonance in Medicine, 2010
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Initial results on in vivo human coronary MR angiography at 7 TMagnetic Resonance in Medicine, 2009
- Simple RF design for human functional and morphological cardiac imaging at 7teslaJournal of Magnetic Resonance, 2009
- Feasibility of Cardiac Gating Free of Interference With Electro-Magnetic Fields at 1.5 Tesla, 3.0 Tesla and 7.0 Tesla Using an MR-StethoscopeInvestigative Radiology, 2009
- B interferometry for the calibration of RF transmitter arraysMagnetic Resonance in Medicine, 2009
- Initial results of cardiac imaging at 7 teslaMagnetic Resonance in Medicine, 2008
- Whole‐body imaging at 7T: Preliminary resultsMagnetic Resonance in Medicine, 2008
- Acoustic Method for Synchronization of Magnetic Resonance Imaging (MRI)Acta Acustica united with Acustica, 2008
- Experimental analysis of parallel excitation using dedicated coil setups and simultaneous RF transmission on multiple channelsMagnetic Resonance in Medicine, 2005