Abstract
Ionization chambers used for reference dosimetry require a local secondary standard ionization chamber with a 60Co absorbed dose to water calibration coefficient N(D,W)(60Co) traceable to a national primary standards dosimetry laboratory or an accredited secondary dosimetry calibration laboratory. Clinic based (in-house) transfer of this coefficient to tertiary reference ionization chambers has traditionally been accomplished with chamber cross calibration in water using a 60Co beam; however, access to 60Co teletherapy machines has become increasingly limited for clinic based physicists. In this work, the accuracy of alternative methods of transferring the N(D,W)(60Co) calibration coefficient using 6 and 18 MV photon beams from a linear accelerator in lieu of 60Co has been investigated for five different setups and four commonly used chamber types.