Image Quality and Radiation Dose of Contrast-Enhanced Chest-CT Acquired on a Clinical Photon-Counting Detector CT vs. Second-Generation Dual-Source CT in an Oncologic Cohort: Preliminary Results
Authors:
Open Access
- 3 June 2022
- journal article
- research article
- Published by MDPI AG in Tomography
- Vol. 8 (3) , 1466-1476
Abstract
Our aim was to compare the image quality and patient dose of contrast-enhanced oncologic chest-CT of a first-generation photon-counting detector (PCD-CT) and a second-generation dual-source dual-energy CT (DSCT). For this reason, one hundred consecutive oncologic patients (63 male, 65 ± 11 years, BMI: 16–42 kg/m2) were prospectively enrolled and evaluated. Clinically indicated contrast-enhanced chest-CT were obtained with PCD-CT and compared to previously obtained chest-DSCT in the same individuals. The median time interval between the scans was three months. The same contrast media protocol was used for both scans. PCD-CT was performed in QuantumPlus mode (obtaining full spectral information) at 120 kVp. DSCT was performed using 100 kV for Tube A and 140 kV for Tube B. “T3D” PCD-CT images were evaluated, which emulate conventional 120 keV polychromatic images. For DSCT, the convolution algorithm was set at I31f with class 1 iterative reconstruction, whereas comparable Br40 kernel and iterative reconstruction strengths (Q1 and Q3) were applied for PCD-CT. Two radiologists assessed image quality using a five-point Likert scale and performed measurements of vessels and lung parenchyma for signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and in the case of pulmonary metastases tumor-to-lung parenchyma contrast ratio. PCD-CT CNRvessel was significantly higher than DSCT CNRvessel (all, p < 0.05). Readers rated image contrast of mediastinum, vessels, and lung parenchyma significantly higher in PCD-CT than DSCT images (p < 0.001). Q3 PCD-CT CNRlung_parenchyma was significantly higher than DSCT CNRlung_parenchyma and Q1 PCD-CT CNRlung_parenchyma (p < 0.01). The tumor-to-lung parenchyma contrast ratio was significantly higher on PCD-CT than DSCT images (0.08 ± 0.04 vs. 0.03 ± 0.02, p < 0.001). CTDI, DLP, SSDE mean values for PCD-CT and DSCT were 4.17 ± 1.29 mGy vs. 7.21 ± 0.49 mGy, 151.01 ± 48.56 mGy * cm vs. 288.64 ± 31.17 mGy * cm and 4.23 ± 0.97 vs. 7.48 ± 1.09, respectively. PCD-CT enables oncologic chest-CT with a significantly reduced dose while maintaining image quality similar to a second-generation DSCT for comparable protocol settings.Keywords
Funding Information
- Baden-Württemberg Ministry of Economic Affairs, Labor and Tourism as part of the “Forum Gesundheitsstandort Baden-Württemberg (35-4223.10/20)
This publication has 26 references indexed in Scilit:
- Single- and dual-energy CT of the abdomen: comparison of radiation dose and image quality of 2nd and 3rd generation dual-source CTEuropean Radiology, 2016
- Scanner thoracique ultra-basse dose : la mort de la radiographie thoracique ?La Presse Médicale, 2016
- To bin or not to bin? The effect of CT system limiting resolution on noise and detectabilityPhysics in Medicine & Biology, 2013
- Size-specific Dose Estimates for Adult Patients at CT of the TorsoRadiology, 2012
- Optimal Tube Potential for Radiation Dose Reduction in Pediatric CT: Principles, Clinical Implementations, and PitfallsRadioGraphics, 2011
- Radiation Dose Reduction in Chest CT: A ReviewAmerican Journal of Roentgenology, 2008
- Angular on‐line tube current modulation in multidetector CT examinations of children and adults: The influence of different scanning parameters on dose reductionMedical Physics, 2007
- CT Dose Reduction and Dose Management Tools: Overview of Available OptionsRadioGraphics, 2006
- Strategies for CT Radiation Dose OptimizationRadiology, 2004
- Technique Factors and Image Quality as Functions of Patient Weight at Abdominal CTRadiology, 2000