Co-registration of cone beam CT and preoperative MRI for improved accuracy of electrode localization following cochlear implantation
- 18 January 2018
- journal article
- research article
- Published by Taylor & Francis in Cochlear Implants International
- Vol. 19 (3) , 147-152
Abstract
Objectives: To investigate the clinical usefulness and practicality of co-registration of Cone Beam CT (CBCT) with preoperative Magnetic Resonance Imaging (MRI) for intracochlear localization of electrodes after cochlear implantation. Methods: Images of 20 adult patients who underwent CBCT after implantation were co-registered with preoperative MRI scans. Time taken for co-registration was recorded. The images were analysed by clinicians of varying levels of expertise to determine electrode position and ease of interpretation. Results: After a short learning curve, the average co-registration time was 10.78 minutes (StdDev 2.37). All clinicians found the co-registered images easier to interpret than CBCT alone. The mean concordance of CBCT vs. co-registered image analysis between consultant otologists was 60% (17-100%) and 86% (60-100%), respectively. The sensitivity and specificity for CBCT to identify Scala Vestibuli insertion or translocation was 100 and 75%, respectively. The negative predictive value was 100%. Discussion: CBCT should be performed following adult cochlear implantation for audit and quality control of surgical technique. If SV insertion or translocation is suspected, co-registration with preoperative MRI should be performed to enable easier analysis. There will be a learning curve for this process in terms of both the co-registration and the interpretation of images by clinicians.Keywords
This publication has 16 references indexed in Scilit:
- The use of cone beam imaging to determine cochlear implant electrode position in human temporal bonesCochlear Implants International, 2013
- Reliability of cone beam computed tomography in scalar localization of the electrode array: a radio histological studyEuropean Archives of Oto-Rhino-Laryngology, 2013
- Midmodiolar Reconstruction as a Valuable Tool to Determine the Exact Position of the Cochlear Implant Electrode ArrayOtology & Neurotology, 2011
- Artifacts of the electrode in cochlea implantation and limits in analysis of deep insertion in cone beam tomography (CBT)European Archives of Oto-Rhino-Laryngology, 2011
- Anatomic verification of a novel method for precise intrascalar localization of cochlear implant electrodes in adult temporal bones using clinically available computed tomographyThe Laryngoscope, 2010
- Radiation dose saving through the use of cone-beam CT in hearing-impaired patientsLa radiologia medica, 2009
- Role of Electrode Placement as a Contributor to Variability in Cochlear Implant OutcomesOtology & Neurotology, 2008
- Intracochlear assessment of electrode position after cochlear implant surgery by means of multislice computer tomography2007
- Quality Control After Insertion of the Nucleus Contour and Contour Advance Electrode in AdultsEar & Hearing, 2007
- Imaging procedures in cochlear implant patients – evaluation of different radiological techniquesActa Oto-Laryngologica, 2004