Harmonization of fibrinogen assay results: study within the framework of the Dutch project ‘Calibration 2000’
- 23 August 2009
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 31 (5) , 513-520
Abstract
In a Dutch project for harmonization of fibrinogen assays, the commutability of potential calibrators for fibrinogen was assessed by means of a twin-study design, which is, in essence, a multicentre, split-patient sample, between-field-methods protocol. The study consisted of simultaneous analysis of fresh-frozen patient plasmas and three potential calibrators for fibrinogen by 48 Dutch laboratories forming 24 couples. The state-of-the-art intralaboratory standard deviation was used to assess the commutability of the potential calibrators. The potential calibrators were commutable for the Clauss, but not for the prothrombin time (PT)-derived assays. One potential calibrator was used in an attempt to harmonize fibrinogen assay results in a Dutch field study. The interlaboratory coefficient of variation (CV) of three out of four test samples could be reduced significantly using the common calibrator. The average overall CV for the four test samples was 10.3% using the routine measurements and 7.8% using the common calibrator. Despite the reduction in the overall CV, the bias between Clauss and PT-derived assay results in two coumarin test samples could not be eliminated.Keywords
This publication has 14 references indexed in Scilit:
- Harmonisation of factor VIII:C assay results: study within the framework of the Dutch project ‘Calibration 2000’British Journal of Haematology, 2005
- Plasma Fibrinogen Level and the Risk of Major Cardiovascular Diseases and Nonvascular MortalityJAMA, 2005
- Japanese collaborative study for fibrinogen assay: variability of the fibrinogen assay between different laboratories does not improve when a common calibrator is usedInternational Journal of Laboratory Hematology, 2005
- Are Results of Fibrinogen Measurements Transferable?Clinical Chemistry and Laboratory Medicine (CCLM), 2003
- Specimen materials, target values and commutability for external quality assessment (proficiency testing) schemesClinica Chimica Acta, 2003
- Commutability Assessment of Potential Reference Materials Using a Multicenter Split-Patient-Sample Between-Field-Methods (Twin-Study) Design: Study within the Framework of the Dutch Project “Calibration 2000”Clinical Chemistry, 2002
- Measurement of plasma fibrinogen concentration by the prothrombin-time-derived methodBlood Coagulation & Fibrinolysis, 1998
- Results of a collaborative study for fibrinogen measurement. Evidence that the use of a common calibrator improves interlaboratory agreementBlood Coagulation & Fibrinolysis, 1994
- Method for the determination of functional (clottable) fibrinogen by the new family of ACL coagulometersThrombosis Research, 1988
- Gerinnungsphysiologische Schnellmethode zur Bestimmung des FibrinogensActa Haematologica, 1957