Wavelength-spatial multiplexed optical coherence tomography via parallel spatial encoding for high-speed, high-precision imaging

Authors:

Abstract
Traditional swept-source optical coherence tomography (SS-OCT) typically rely on point-by-point scanning and complex signal synchronization mechanisms, while their imaging depth and measurement range are fundamentally constrained by the limited resampling rate of k-clock signals. To address this, we propose wavelength-spatial multiplexed OCT with parallel spatial encoding for high-speed and high-precision imaging. A diffraction grating is employed to convert conventional point-scanning measurement into line-scanning measurement by establishing a wavelength-to-space mapping, allowing the line spatial information of the target to be encoded in parallel within a single interference sweep. To recover the encoded line spatial information, an equal-frequency subdivision resampling (EFSR) method is proposed to resample the interference signal on a refined grid, thereby compensating for errors caused by the nonlinear sweep of the light source. The resampled signal is then processed using short-time Fourier transform (STFT) analysis to reconstruct depth-resolved profiles along the encoded line. Experimental results demonstrate a lateral measurement error below 0.90 μm. For axial measurements, the standard deviation across 100 repeated measurements is 117.6 nm. The system also clearly resolves multilayer interfaces in tomographic images. The proposed approach combines parallel spatial encoding with k-clock-based resampling to enable high-precision three-dimensional imaging.

This publication has 18 references indexed in Scilit: