Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI‐Linac prototypes
- 10 December 2017
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 45 (1) , 479-487
Abstract
Purpose: Longitudinal magnetic fields narrow beam penumbra and tighten lateral spread of secondary electrons in air cavities, including lung tissue. Gafchromic® EBT3 film was used to investigate differences between penumbra in solid water and solid lung, without a magnetic field (0 T) and with two field strengths (0.9 and 1.5 T).Methods: The first prototype of the Australian MRI‐linac consisted of a 1.5 T Siemens Sonata MRI and Varian industrial linatron (nominal 4 MV). The second prototype replaced the Sonata with a 1.0 T Agilent split‐bore magnet. Measurements were completed at 0.9 T to maintain the same source‐to‐surface distance between set‐ups. Gammex‐rmi® solid water with 50 mm of CIRS solid lung inserted as a lung cavity was positioned inside each magnet. This was compared to the same set‐up with solid water only, where film measurements were completed at solid water equivalent depths corresponding to entrance interface/mid/exit interface positions of solid lung from the first set‐up. Multileaf collimator (MLC)‐defined field sizes were set to 3 × 3 cm2 and 10 × 10 cm2. The 80%–20% penumbral width was determined.Results: Under 1.5 T conditions, penumbra narrowing occurred up to 4.4 ± 0.1 mm compared to 0 T. As expected, the effect was less for 0.9 T, which resulted in a maximum narrowing of 2.5 ± 0.1 mm. Exit profile penumbra were more affected than entrance penumbra by up to 2.6 ± 0.2 mm. The 1.5 T field brought the solid water and lung penumbral widths more into alignment by a maximum difference of 0.4 ± 0.1 mm.Conclusions: The trimming of penumbral widths due to magnetic fields in solid water and lung was demonstrated and compared to 0 T. The 0.9 and 1.5 T field trimmed the penumbra by up to 2.5 ± 0.1 mm and 4.4 ± 0.1 mm respectively.Keywords
Funding Information
- National Health and Medical Research Council (APP1036078)
- Australian Cancer Research Foundation
This publication has 17 references indexed in Scilit:
- Technical Note: Experimental results from a prototype high‐field inline MRI‐linacMedical Physics, 2016
- The Australian Magnetic Resonance Imaging–Linac ProgramSeminars in Radiation Oncology, 2014
- The Rotating Biplanar Linac–Magnetic Resonance Imaging SystemSeminars in Radiation Oncology, 2014
- The Magnetic Resonance Imaging–Linac SystemSeminars in Radiation Oncology, 2014
- The ViewRay System: Magnetic Resonance–Guided and Controlled RadiotherapySeminars in Radiation Oncology, 2014
- Electron contamination modeling and skin dose in 6 MV longitudinal field MRIgRT: Impact of the MRI and MRI fringe fieldMedical Physics, 2012
- Lung dosimetry in a linac‐MRI radiotherapy unit with a longitudinal magnetic fieldMedical Physics, 2010
- Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strengthPhysics in Medicine & Biology, 2008
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beamsMedical Physics, 1993