Sarah J. Alnaghy
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Research Subjects
Sarah J. Alnaghy has a total of 28 co-authors. Here are the top co-authors by number of shared publications:
- Peter Metcalfe
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, Australia
5 shared publications Jarrad Begg
Ingham Institute for Applied Medical Research, Liverpool, NSW, 2170, Australia
4 shared publicationsTrent Causer
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, 2522, Australia
4 shared publicationsBin Dong
Ingham Institute for Applied Medical Research, Liverpool, NSW, 2170, Australia
3 shared publicationsLois Holloway
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, 2522, Australia
3 shared publicationsMaegan Gargett
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW, 2522, Australia
3 shared publications
Latest Publications
- 9 January 2020
- Published by Wiley in Medical Physics
- Vol. 47 (4) , 1920-1929
High resolution silicon array detector implementation in an inline MRI‐linac
Purpose: Dynamic dosimaging is a concept whereby a detector in motion is tracked with magnetic resonance imaging (MRI) to validate the amount and position of dose in a radiation therapy treatment on an MRI‐linac. This work takes steps toward the realization of dynamic dosimaging with the novel high resolution silicon array detector: MagicPlate‐512 (M512). The performance of the M512 was assessed in a 1.0 T inline MRI‐linac, without simultaneous imaging and then during an imaging sequence, both during dosimetry. MR images were acquired to determine the effect of the detector and its components on image quality.Methods: Beam profiles were measured using the M512 on the Australian MRI‐Linac and a comparison made with Gafchromic EBT3 film to investigate any intrinsic magnetic field effects in the silicon. The M512 has 512 sensitive volumes, each 0.5 × 0.5 × 0.037 mm3 in dimension, organized in a two‐dimensional array. Small field sizes up to 4.2 × 3.8 cm2 were investigated in both solid water and then solid lung phantoms. Beam profiles taken at 1.0 T were compared to 0 T conditions, and also to profiles taken during a gradient echo (GRE) imaging sequence. Differences in 80%‐20% penumbral width and full width at half maximum (FWHM) were investigated. Localizer MR images were acquired of the detector adjacent to a water phantom.Results: Good agreement was observed between the M512 and film, with average differences in penumbral width and FWHM of <1 mm in the absence of the imaging sequence. Concurrent imaging widened the penumbra by up to 1.2 mm due to RF noise affecting the detector; film profiles were unchanged. Magnetic resonance images were affected by noise, in particular, due to the large amount of aluminum present, as well as from the USB cable, which acted as an antenna. Unfortunately, due to these issues, suitable dynamic dose imaging was not achieved with the current M512/phantom configuration and the MRI‐linac. However, progress was made toward achieving this goal for future work.Conclusions: The M512 silicon array detector successfully measured high‐resolution beam profiles in agreement with Gafchromic film to within an average of <1 mm on the first MRI‐linac in Australia. More effective noise reduction will be required for the achievement of dynamic dosimaging in the future.
- 9 January 2020
- Published by Wiley in Medical Physics
- Vol. 47 (4) , 1920-1929
High resolution silicon array detector implementation in an inline MRI‐linac
Purpose: Dynamic dosimaging is a concept whereby a detector in motion is tracked with magnetic resonance imaging (MRI) to validate the amount and position of dose in a radiation therapy treatment on an MRI‐linac. This work takes steps toward the realization of dynamic dosimaging with the novel high resolution silicon array detector: MagicPlate‐512 (M512). The performance of the M512 was assessed in a 1.0 T inline MRI‐linac, without simultaneous imaging and then during an imaging sequence, both during dosimetry. MR images were acquired to determine the effect of the detector and its components on image quality.Methods: Beam profiles were measured using the M512 on the Australian MRI‐Linac and a comparison made with Gafchromic EBT3 film to investigate any intrinsic magnetic field effects in the silicon. The M512 has 512 sensitive volumes, each 0.5 × 0.5 × 0.037 mm3 in dimension, organized in a two‐dimensional array. Small field sizes up to 4.2 × 3.8 cm2 were investigated in both solid water and then solid lung phantoms. Beam profiles taken at 1.0 T were compared to 0 T conditions, and also to profiles taken during a gradient echo (GRE) imaging sequence. Differences in 80%‐20% penumbral width and full width at half maximum (FWHM) were investigated. Localizer MR images were acquired of the detector adjacent to a water phantom.Results: Good agreement was observed between the M512 and film, with average differences in penumbral width and FWHM of <1 mm in the absence of the imaging sequence. Concurrent imaging widened the penumbra by up to 1.2 mm due to RF noise affecting the detector; film profiles were unchanged. Magnetic resonance images were affected by noise, in particular, due to the large amount of aluminum present, as well as from the USB cable, which acted as an antenna. Unfortunately, due to these issues, suitable dynamic dose imaging was not achieved with the current M512/phantom configuration and the MRI‐linac. However, progress was made toward achieving this goal for future work.Conclusions: The M512 silicon array detector successfully measured high‐resolution beam profiles in agreement with Gafchromic film to within an average of <1 mm on the first MRI‐linac in Australia. More effective noise reduction will be required for the achievement of dynamic dosimaging in the future.
- 27 June 2019
- Published by Wiley in Medical Physics
- Vol. 46 (9) , 4224-4232
A feasibility study for high‐resolution silicon array detector performance in the magnetic field of a permanent magnet system
Purpose: Magnetic field effects on dose distribution and detector functionality must be well understood. The detector utilized to investigate these magnetic field effects was the DUO silicon array detector; the performance of this high spatial resolution detector was assessed under these conditions. The results were compared to Gafchromic EBT3 film to highlight any intrinsic magnetic field effects in the silicon. The results were also compared to previously published MagicPlate‐512 (M512) data. The DUO has an improved spatial resolution (200 µm) over the M512 (2 mm).Methods: A permanent magnet named Magnetic Apparatus for RaDiation Oncology Studies (MARDOS) paired with a standard linear accelerator (linac) enables either transverse (1.2 T) or inline (0.95 T) orientations of the magnetic field with respect to the radiation beam. A 6 MV Varian 2100C Linac provided the radiation component for the measurements. The DUO detector has 505 sensitive volumes (each volume measuring 800 × 40 × 100 µm3) organized in two orthogonal, linear arrays. The DUO was embedded in a solid water phantom in the first set‐up and then a solid lung phantom in the second set‐up and placed between the magnet cones. Beam profiles were compared under the magnetic field conditions and 0 T. Small field sizes from 0.8 × 0.8 cm2up to 2.3 × 2.3 cm2were investigated. The size of the air gap above the sensitive volumes of the DUO was investigated in the transverse orientation to assess the anticipated magnetic field effects. Full width at half maximum (FWHM), 80–20% penumbral widths and maximum dose differences between detectors and between the presence/absence of a magnetic field were investigated. Symmetry was also assessed for investigation of profile skewness under the transverse field.Results: The penumbral widths measured by the DUO detector demonstrated good agreement with film and the M512 to within an average of 0.5 mm (within uncertainty: ±1 mm). The static inline magnetic field had minimal effect on the profiles in solid water. As expected, the lower density of solid lung meant that this material was more susceptible to demonstrating magnetic field effects in the dose deposited. The greatest penumbral narrowing due to the inline field (0.7 mm) occurred in lung. Central axis dose increase was greatest in lung (maximum: 9%). The transverse field widened penumbra, most notably in the solid lung phantom, by a maximum of 2.3 mm. The largest asymmetry due to the transverse field (4.6%) was also in solid lung. When the air gap above the DUO was filled with bolus, the dose maximum measured by the DUO was within 1.4% of film.Conclusions: The DUO detector has been shown to be successful in accurately describing the dose changes for small field sizes to within a 200‐µm resolution in an environment resembling that of an MRI‐linac. The DUO measurements were in agreement with both film and the M512 measurements, and therefore the DUO was found to be an appropriate alternative to the M512, with improvement in terms of its higher spatial resolution. MARDOS provided a suitable environment for these preliminary tests before progressing to the MRI‐linac.
- 27 June 2019
- Published by Wiley in Medical Physics
- Vol. 46 (9) , 4224-4232
A feasibility study for high‐resolution silicon array detector performance in the magnetic field of a permanent magnet system
Purpose: Magnetic field effects on dose distribution and detector functionality must be well understood. The detector utilized to investigate these magnetic field effects was the DUO silicon array detector; the performance of this high spatial resolution detector was assessed under these conditions. The results were compared to Gafchromic EBT3 film to highlight any intrinsic magnetic field effects in the silicon. The results were also compared to previously published MagicPlate‐512 (M512) data. The DUO has an improved spatial resolution (200 µm) over the M512 (2 mm).Methods: A permanent magnet named Magnetic Apparatus for RaDiation Oncology Studies (MARDOS) paired with a standard linear accelerator (linac) enables either transverse (1.2 T) or inline (0.95 T) orientations of the magnetic field with respect to the radiation beam. A 6 MV Varian 2100C Linac provided the radiation component for the measurements. The DUO detector has 505 sensitive volumes (each volume measuring 800 × 40 × 100 µm3) organized in two orthogonal, linear arrays. The DUO was embedded in a solid water phantom in the first set‐up and then a solid lung phantom in the second set‐up and placed between the magnet cones. Beam profiles were compared under the magnetic field conditions and 0 T. Small field sizes from 0.8 × 0.8 cm2up to 2.3 × 2.3 cm2were investigated. The size of the air gap above the sensitive volumes of the DUO was investigated in the transverse orientation to assess the anticipated magnetic field effects. Full width at half maximum (FWHM), 80–20% penumbral widths and maximum dose differences between detectors and between the presence/absence of a magnetic field were investigated. Symmetry was also assessed for investigation of profile skewness under the transverse field.Results: The penumbral widths measured by the DUO detector demonstrated good agreement with film and the M512 to within an average of 0.5 mm (within uncertainty: ±1 mm). The static inline magnetic field had minimal effect on the profiles in solid water. As expected, the lower density of solid lung meant that this material was more susceptible to demonstrating magnetic field effects in the dose deposited. The greatest penumbral narrowing due to the inline field (0.7 mm) occurred in lung. Central axis dose increase was greatest in lung (maximum: 9%). The transverse field widened penumbra, most notably in the solid lung phantom, by a maximum of 2.3 mm. The largest asymmetry due to the transverse field (4.6%) was also in solid lung. When the air gap above the DUO was filled with bolus, the dose maximum measured by the DUO was within 1.4% of film.Conclusions: The DUO detector has been shown to be successful in accurately describing the dose changes for small field sizes to within a 200‐µm resolution in an environment resembling that of an MRI‐linac. The DUO measurements were in agreement with both film and the M512 measurements, and therefore the DUO was found to be an appropriate alternative to the M512, with improvement in terms of its higher spatial resolution. MARDOS provided a suitable environment for these preliminary tests before progressing to the MRI‐linac.
- Open Access
- 4 September 2018
- Published by Wiley in Medical Physics
- Vol. 45 (10) , 4783-4787
Erratum: “Technical Note: Penumbral Width Trimming in Solid Lung Dose Profiles for 0.9 T and 1.5 T MRI‐Linac Prototypes” [Med. Phys. 45(1), 479‐487 (2018)]
- 4 September 2018
- Published by Wiley in Medical Physics
- Vol. 45 (10) , 4783-4787
Open AccessErratum: “Technical Note: Penumbral Width Trimming in Solid Lung Dose Profiles for 0.9 T and 1.5 T MRI‐Linac Prototypes” [Med. Phys. 45(1), 479‐487 (2018)]
- 10 December 2017
- Published by Wiley in Medical Physics
- Vol. 45 (1) , 479-487
Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI‐Linac prototypes
Purpose: Longitudinal magnetic fields narrow beam penumbra and tighten lateral spread of secondary electrons in air cavities, including lung tissue. Gafchromic® EBT3 film was used to investigate differences between penumbra in solid water and solid lung, without a magnetic field (0 T) and with two field strengths (0.9 and 1.5 T).Methods: The first prototype of the Australian MRI‐linac consisted of a 1.5 T Siemens Sonata MRI and Varian industrial linatron (nominal 4 MV). The second prototype replaced the Sonata with a 1.0 T Agilent split‐bore magnet. Measurements were completed at 0.9 T to maintain the same source‐to‐surface distance between set‐ups. Gammex‐rmi® solid water with 50 mm of CIRS solid lung inserted as a lung cavity was positioned inside each magnet. This was compared to the same set‐up with solid water only, where film measurements were completed at solid water equivalent depths corresponding to entrance interface/mid/exit interface positions of solid lung from the first set‐up. Multileaf collimator (MLC)‐defined field sizes were set to 3 × 3 cm2 and 10 × 10 cm2. The 80%–20% penumbral width was determined.Results: Under 1.5 T conditions, penumbra narrowing occurred up to 4.4 ± 0.1 mm compared to 0 T. As expected, the effect was less for 0.9 T, which resulted in a maximum narrowing of 2.5 ± 0.1 mm. Exit profile penumbra were more affected than entrance penumbra by up to 2.6 ± 0.2 mm. The 1.5 T field brought the solid water and lung penumbral widths more into alignment by a maximum difference of 0.4 ± 0.1 mm.Conclusions: The trimming of penumbral widths due to magnetic fields in solid water and lung was demonstrated and compared to 0 T. The 0.9 and 1.5 T field trimmed the penumbra by up to 2.5 ± 0.1 mm and 4.4 ± 0.1 mm respectively.
- 10 December 2017
- Published by Wiley in Medical Physics
- Vol. 45 (1) , 479-487
Technical Note: Penumbral width trimming in solid lung dose profiles for 0.9 and 1.5 T MRI‐Linac prototypes
Purpose: Longitudinal magnetic fields narrow beam penumbra and tighten lateral spread of secondary electrons in air cavities, including lung tissue. Gafchromic® EBT3 film was used to investigate differences between penumbra in solid water and solid lung, without a magnetic field (0 T) and with two field strengths (0.9 and 1.5 T).Methods: The first prototype of the Australian MRI‐linac consisted of a 1.5 T Siemens Sonata MRI and Varian industrial linatron (nominal 4 MV). The second prototype replaced the Sonata with a 1.0 T Agilent split‐bore magnet. Measurements were completed at 0.9 T to maintain the same source‐to‐surface distance between set‐ups. Gammex‐rmi® solid water with 50 mm of CIRS solid lung inserted as a lung cavity was positioned inside each magnet. This was compared to the same set‐up with solid water only, where film measurements were completed at solid water equivalent depths corresponding to entrance interface/mid/exit interface positions of solid lung from the first set‐up. Multileaf collimator (MLC)‐defined field sizes were set to 3 × 3 cm2 and 10 × 10 cm2. The 80%–20% penumbral width was determined.Results: Under 1.5 T conditions, penumbra narrowing occurred up to 4.4 ± 0.1 mm compared to 0 T. As expected, the effect was less for 0.9 T, which resulted in a maximum narrowing of 2.5 ± 0.1 mm. Exit profile penumbra were more affected than entrance penumbra by up to 2.6 ± 0.2 mm. The 1.5 T field brought the solid water and lung penumbral widths more into alignment by a maximum difference of 0.4 ± 0.1 mm.Conclusions: The trimming of penumbral widths due to magnetic fields in solid water and lung was demonstrated and compared to 0 T. The 0.9 and 1.5 T field trimmed the penumbra by up to 2.5 ± 0.1 mm and 4.4 ± 0.1 mm respectively.
- 4 November 2016
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 39 (4) , 921-932
Initial experiments with gel-water: towards MRI-linac dosimetry and imaging
Tracking the position of a moving radiation detector in time and space during data acquisition can replicate 4D image-guided radiotherapy (4DIGRT). Magnetic resonance imaging (MRI)-linacs need MRI-visible detectors to achieve this, however, imaging solid phantoms is an issue. Hence, gel-water, a material that provides signal for MRI-visibility, and which will in future work, replace solid water for an MRI-linac 4DIGRT quality assurance tool, is discussed. MR and CT images of gel-water were acquired for visualisation and electron density verification. Characterisation of gel-water at 0 T was compared to Gammex-RMI solid water, using MagicPlate-512 (M512) and RMI Attix chamber; this included percentage depth dose, tissue-phantom ratio (TPR20/10), tissue-maximum ratio (TMR), profiles, output factors, and a gamma analysis to investigate field penumbral differences. MR images of a non-powered detector in gel-water demonstrated detector visualisation. The CT-determined gel-water electron density agreed with the calculated value of 1.01. Gel-water depth dose data demonstrated a maximum deviation of 0.7% from solid water for M512 and 2.4% for the Attix chamber, and by 2.1% for TPR20/10 and 1.0% for TMR. FWHM and output factor differences between materials were ≤0.3 and ≤1.4%. M512 data passed gamma analysis with 100% within 2%, 2 mm tolerance for multileaf collimator defined fields. Gel-water was shown to be tissue-equivalent for dosimetry and a feasible option to replace solid water.
- 4 November 2016
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 39 (4) , 921-932
Initial experiments with gel-water: towards MRI-linac dosimetry and imaging
Tracking the position of a moving radiation detector in time and space during data acquisition can replicate 4D image-guided radiotherapy (4DIGRT). Magnetic resonance imaging (MRI)-linacs need MRI-visible detectors to achieve this, however, imaging solid phantoms is an issue. Hence, gel-water, a material that provides signal for MRI-visibility, and which will in future work, replace solid water for an MRI-linac 4DIGRT quality assurance tool, is discussed. MR and CT images of gel-water were acquired for visualisation and electron density verification. Characterisation of gel-water at 0 T was compared to Gammex-RMI solid water, using MagicPlate-512 (M512) and RMI Attix chamber; this included percentage depth dose, tissue-phantom ratio (TPR20/10), tissue-maximum ratio (TMR), profiles, output factors, and a gamma analysis to investigate field penumbral differences. MR images of a non-powered detector in gel-water demonstrated detector visualisation. The CT-determined gel-water electron density agreed with the calculated value of 1.01. Gel-water depth dose data demonstrated a maximum deviation of 0.7% from solid water for M512 and 2.4% for the Attix chamber, and by 2.1% for TPR20/10 and 1.0% for TMR. FWHM and output factor differences between materials were ≤0.3 and ≤1.4%. M512 data passed gamma analysis with 100% within 2%, 2 mm tolerance for multileaf collimator defined fields. Gel-water was shown to be tissue-equivalent for dosimetry and a feasible option to replace solid water.
- Open Access
- 8 May 2015
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (3) , 107-117
In vivo endorectal dosimetry of prostate tomotherapy using dual MOSkin detectors
Verification of dose to the anterior rectal wall in helical tomotherapy to the prostate is important due to the close proximity of the rectal wall to the treatment field. The steep dose gradient makes these measurements challenging. A phantom‐based study was completed, aimed at developing a system for measurement of anterior rectal wall doses during hypofractionated prostate stereotactic body radiotherapy (SBRT) utilizing tomotherapy delivery. An array of four dual MOSkin™ dosimeters, spaced 1 cm apart, was placed on a replica Rectafix® immobilization spacer device. This Perspex probe is a more rigid alternative to rectal balloons, to improve geometric reproducibility. The doses at each point were measured in real time and compared to doses calculated by the treatment planning system (TPS). Additionally, distance‐to‐agreement (DTA) measurements were acquired to assist in the comparison of measured and predicted doses. All dual MOSkin detectors measured dose to within of the TPS at the anterior rectal wall. Whilst several points were outside of experimental error, the largest deviation from the TPS predicted dose represented a DTA of only 1.3 mm, within the acceptable DTA tolerance of 3 mm. Larger deviations of up to −11.9% were observed for the posterior and side walls; however, if acceptable DTA measurements are accounted for, then an agreement of 75% was observed. Although larger differences were observed at the other rectal wall locations, the overall effect of dose at these points was not as significant, given the lower doses. Despite the very high‐dose gradient region, real‐time measurements of the anterior rectal wall doses were within acceptable limits of TPS‐predicted doses. The differences between measured and planned data were due to difficulties in precisely locating each detector on the TPS dose grid, which presented large variations in dose between CT voxels in regions of steep dose gradients. The dual MOSkin system would, therefore, be a useful device for detecting errors in real time, such as patient shifts or incorrect setup, during tomotherapy of the prostate.PACS numbers: 87.53.Ly, 87.55.km, 87.55.N‐
- 8 May 2015
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (3) , 107-117
Open AccessIn vivo endorectal dosimetry of prostate tomotherapy using dual MOSkin detectors
Verification of dose to the anterior rectal wall in helical tomotherapy to the prostate is important due to the close proximity of the rectal wall to the treatment field. The steep dose gradient makes these measurements challenging. A phantom‐based study was completed, aimed at developing a system for measurement of anterior rectal wall doses during hypofractionated prostate stereotactic body radiotherapy (SBRT) utilizing tomotherapy delivery. An array of four dual MOSkin™ dosimeters, spaced 1 cm apart, was placed on a replica Rectafix® immobilization spacer device. This Perspex probe is a more rigid alternative to rectal balloons, to improve geometric reproducibility. The doses at each point were measured in real time and compared to doses calculated by the treatment planning system (TPS). Additionally, distance‐to‐agreement (DTA) measurements were acquired to assist in the comparison of measured and predicted doses. All dual MOSkin detectors measured dose to within of the TPS at the anterior rectal wall. Whilst several points were outside of experimental error, the largest deviation from the TPS predicted dose represented a DTA of only 1.3 mm, within the acceptable DTA tolerance of 3 mm. Larger deviations of up to −11.9% were observed for the posterior and side walls; however, if acceptable DTA measurements are accounted for, then an agreement of 75% was observed. Although larger differences were observed at the other rectal wall locations, the overall effect of dose at these points was not as significant, given the lower doses. Despite the very high‐dose gradient region, real‐time measurements of the anterior rectal wall doses were within acceptable limits of TPS‐predicted doses. The differences between measured and planned data were due to difficulties in precisely locating each detector on the TPS dose grid, which presented large variations in dose between CT voxels in regions of steep dose gradients. The dual MOSkin system would, therefore, be a useful device for detecting errors in real time, such as patient shifts or incorrect setup, during tomotherapy of the prostate.PACS numbers: 87.53.Ly, 87.55.km, 87.55.N‐