Thermal Cycling Effect on Locator System Retention and Metal Surface Roughness
Open Access
- 21 January 2022
- journal article
- research article
- Published by Wiley in Journal of Prosthodontics
- Vol. 31 (9) , 771-777
Abstract
Purpose: To estimate the effects of the thermal cycling (TC) process on the metal surfaces of Locators, as well as retention loss, and the correlation between them.Materials and methods: Twenty‐five new Locator R‐Tx were included in the study. Four areas were marked on each Locators’ patrix metal surface and scanned using a confocal scanner (μsurf explorer; NanoFocus). Three surface roughness parameters were measured in the scans: Sa (average distance of peaks from the central plain of the area), Vmp (volume of the peaks in the area), and Spc (mean curvature of the peaks describing the degree of their sharpness). Retention test was performed using Instron® 4500 compression tension tensile tester at a speed of 10 mm/min. The retention tests were done using a working model made of two acrylic blocks in which the Locator system parts were inserted. The surface parameters measurements and the retention tests were performed 2 times, once before and once after TC. The Locators were subjected to 15,000 TC cycles by investing them into 2 tubs with different water temperatures, 55°C and 5°C. During each 60‐second cycle, the Locators were submerged in each tub for 20 seconds, with a 10 second transition time between the tubs. The post‐TC retention and surface parameters measurements were compared with those prior to TC and the prior to TC measurements served as controls. Changes in parameters before and after TC were analyzed by a two‐way ANOVA nested model with random intercept and slope by restricted maximum likelihood method. Correlation between retention and surface parameters was quantified and examined using Kendall's correlation test. The findings were considered statistically significant if p < 0.05.Results: There was a significant decrease in retention of 16.6N at the second retention test (p < 0.001). A significant statistical decrease in surface parameters were measured after TC process, Sa and Vmp (18 ×10–3 μm, p = 0.041 and 0.94 ×10–3 1/μm, p = 0.001, respectively). A significant statistical increase in Spc of 6.4 ×10–3 μm3/μm2 (p = 0.023) was noticed. The correlation between retention decreases and surface changes was not statistically significant.Conclusion: The TC process causes a substantial reduction in retention to the Locator system over time. In addition, TC causes significant but minor changes to the Locator surface area. Most of the changes are in the horizontal dimension.Keywords
This publication has 34 references indexed in Scilit:
- Oral Health-related Quality of LifeJournal of Dental Research, 2011
- Wear simulation effects on overdenture stud attachmentsDental Materials Journal, 2011
- Resistance to Wear of Four Matrices with Ball Attachments for Implant Overdentures: A Fatigue StudyJournal of Prosthodontics, 2010
- A Comparison of Three Different Attachment Systems for Mandibular Two‐Implant Overdentures: One‐Year ReportClinical Implant Dentistry and Related Research, 2010
- Attachment Systems for Implant Retained Overdentures: A Literature ReviewImplant Dentistry, 2006
- Effect of Implant Angulation upon Retention of Overdenture AttachmentsJournal of Prosthodontics, 2005
- Retention Characteristics of Attachment Systems for Implant OverdenturesJournal of Prosthodontics, 2004
- Mandibular overdentures supported by two or four endosseous implantsClinical Oral Implants Research, 2004
- The Oxidation and Corrosion Behavior of Nonoxide Ceramic Matrix Composites2000
- The efficiency of cobalt samarium (Co5Sm) magnets as retention units for overdenturesJournal of Dentistry, 1983