1-year outcomes of the Xen45 glaucoma implant
- 14 December 2018
- journal article
- research article
- Published by Springer Nature in Eye
- Vol. 33 (5) , 761-766
Abstract
To describe the 12-month outcomes of the Xen45 glaucoma stent. Non-comparative retrospective study of all cases who underwent Xen glaucoma surgery in April 2017 or earlier and completed 12 months of follow-up. The primary outcome measures were intraocular pressure (IOP) reduction and number of glaucoma medications at 12 months postoperatively. The secondary outcome measures were surgical complications and the success rate of surgery at 1 year. Success rate was defined according to the multiple IOP thresholds of 15 mmHg, 18 mmHg, and 21 mmHg with all requiring a drop of 20% and no additional glaucoma surgery. Revision or needling of the Xen conjunctival bleb was not considered to constitute a surgical failure. Sixty-eight eyes were included in the study. Mean IOP dropped from 22.1 mmHg preoperatively to 14.8 mmHg at 12 months, a 33% drop (p < 0.0001). Mean number of glaucoma medications reduced from 2.9 preoperatively to 1.1 at 12 months (p < 0.0001). In total, 54.4% of cases were back on glaucoma medications by 12 months. Success rate varied from 32.4% when defined as IOP ≤ 15 mmHg and ≥ 6 mmHg and ≥ 20% reduction without medications to 70.6% when defined as IOP ≤ 21 mmHg and ≥ 6 mmHg and ≥ 20% reduction with or without medications. Thirty cases (44.1%) required bleb needling or surgical revision. The Xen45 is effective at reducing IOP and glaucoma medication use at 12 months postoperatively. Patients considering this procedure should be warned that by 12 months postoperatively there is a significant chance of requiring postoperative bleb intervention and glaucoma drops.Keywords
This publication has 15 references indexed in Scilit:
- Minimally-invasive glaucoma surgeries (MIGS) for open angle glaucoma: A systematic review and meta-analysisPLOS ONE, 2017
- Practice Preferences for Glaucoma Surgery: A Survey of the American Glaucoma SocietyJournal of Glaucoma, 2017
- Evaluation of Bleb Morphology and Reduction in IOP and Glaucoma Medication following Implantation of a Novel Gel StentJournal of Ophthalmology, 2017
- The Effects of Phacoemulsification on Intraocular Pressure and Topical Medication Use in Patients With Glaucoma: A Systematic Review and Meta-analysis of 3-Year DataJournal of Glaucoma, 2017
- Minimally invasive combined glaucoma and cataract surgery: clinical results of the smallest ab interno gel stentInternational Ophthalmology, 2017
- XEN Glaucoma Implant with Mitomycin C 1-Year Follow-Up: Result and ComplicationsJournal of Ophthalmology, 2017
- Survey of glaucoma surgical preferences and post‐operative care in the United KingdomClinical & Experimental Ophthalmology, 2016
- Cirugía combinada mediante facoemulsificación e implante XEN45 con acceso temporal y 2 únicas incisionesArchivos de la Sociedad Española de Oftalmología, 2016
- Early Clinical Results of a Novel Ab Interno Gel Stent for the Surgical Treatment of Open-angle GlaucomaJournal of Glaucoma, 2016
- Phacoemulsification combined with a new ab interno gel stent to treat open-angle glaucoma: Pilot studyJournal of Cataract & Refractive Surgery, 2015