Combined Use of a Transmission Detector and an EPID-Based In Vivo Dose Monitoring System in External Beam Whole Breast Irradiation: A Study with an Anthropomorphic Female Phantom
Authors:
- Chiara ArilliCAChiara Arilli
- Yannik WandaelYWYannik Wandael
- Chiara GaleottiCGChiara Galeotti
- Livia MarrazzoLMLivia Marrazzo
- Silvia CalusiSCSilvia Calusi
- Mattia GrusioMGMattia Grusio
- Isacco DesideriIDIsacco Desideri
- Franco FusiFFFranco Fusi
- Angelo PiermatteiAPAngelo Piermattei
- Stefania PallottaSPStefania Pallotta
- Cinzia TalamontiCTCinzia Talamonti
Open Access
- 28 October 2020
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 10 (21) , 7611
Abstract
We evaluate the combined usage of two systems, the Integral Quality Monitor (IQM) transmission detector and SoftDiso software, for in vivo dose monitoring by simultaneous detection of delivery and patient setup errors in whole breast irradiation. An Alderson RANDO phantom was adapted with silicon breast prostheses to mimic the female anatomy. Plans with simulated delivery errors were created from a reference left breast plan, and patient setup errors were simulated by moving the phantom. Deviations from reference values recorded by both monitoring systems were measured for all plans and phantom positions. A 2D global gamma analysis was performed in SoftDiso for all phantom displacements. Both IQM signals and SoftDiso R-values are sensitive to small MU variations. However, only IQM is sensitive to jaw position variations. Conversely, IQM is unable to detect patient positioning errors, and the R-value has good sensitivity to phantom displacements. A gamma comparison analysis allows one to determine alert thresholds to detect phantom shifts or relatively large rotations. The combined use of the IQM and SoftDiso allows for fast identification of both delivery and setup errors and substantially reduces the impact of error identification and correction on the treatment workflow.Keywords
This publication has 41 references indexed in Scilit:
- Dosimetric comparison of left-sided whole-breast irradiation with 3DCRT, forward-planned IMRT, inverse-planned IMRT, helical tomotherapy, and volumetric arc therapyPhysica Medica, 2015
- Quality Control Quantification (QCQ): A Tool to Measure the Value of Quality Control Checks in Radiation OncologyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Catching errors with in vivo EPID dosimetryMedical Physics, 2010
- Dosimetric comparison of intensity modulated radiotherapy techniques and standard wedged tangents for whole breast radiotherapy*Journal of Medical Imaging and Radiation Oncology, 2009
- A planning comparison of dose patterns in organs at risk and predicted risk for radiation induced malignancy in the contralateral breast following radiation therapy of primary breast using conventional, IMRT and Volumetric modulated arc treatment techniquesActa Oncologica, 2009
- Comparison of conventional inserts and an add-on electron MLC for chest wall irradiation of left-sided breast cancerActa Oncologica, 2009
- Clinical experience in breast irradiation with intensity modulated photon beamsActa Oncologica, 2005
- Twenty-Year Follow-up of a Randomized Trial Comparing Total Mastectomy, Lumpectomy, and Lumpectomy plus Irradiation for the Treatment of Invasive Breast CancerNew England Journal of Medicine, 2002
- Twenty-Year Follow-up of a Randomized Study Comparing Breast-Conserving Surgery with Radical Mastectomy for Early Breast CancerNew England Journal of Medicine, 2002
- Twenty-Five-Year Follow-up of a Randomized Trial Comparing Radical Mastectomy, Total Mastectomy, and Total Mastectomy Followed by IrradiationNew England Journal of Medicine, 2002