Abstract
Aim The MiniMed™ 780G system utilizes an Advanced Hybrid Closed-Loop algorithm, which is optimized to perform best at a glucose target of 100 mg/dL. The system also includes a temporary target feature (TT) of 150 mg/dL that can be set when higher glucose utilization is anticipated (e.g., exercise/strenuous work). Methods Real-world data from 1002 randomly selected users of the MiniMed™ 780G and MiniMed™ 670G systems were used to investigate the TT functionality (n = 501 MiniMed™ 780G users, n = 501 MiniMed™ 670G users). Continuous glucose monitoring (CGM) metrics were calculated for 4-h intervals (4 h prior to commencing TT use and 0–4, 4–8, 8–12, 12–16, 16–20 and 20–24 h after TT initiation). Results CGM results showed that, with both systems, the time-in-range (TIR) during TT periods mostly equalled or exceeded TIR in corresponding non-TT periods. In the first 4 h after TT initiation, the MiniMed™ 780G system demonstrated a TIR and time in tight range (TITR) of 71.6% and 46.4%, respectively, [a 1.6 (p = 0.11) and 2.6 (p = 0.009) percentage point benefit in TIR and TITR, respectively, versus the 670G system]. Mean time-below-range remained below target in both periods. Conclusions These findings confirm that the TT setting performs as intended, allowing users to maintain TIR during strenuous activity without compromising safety.
Funding Information
  • Medtronic

This publication has 13 references indexed in Scilit: