Abstract
Celiac disease is a chronic autoimmune disorder triggered by gluten consumption which affects between 0.5% and 1% of the global population. Current diagnostic methods still require invasive procedures such as intestinal biopsy. LMR-based sensors hold great potential for the development of reliable and user-friendly devices for diagnosing this condition. In this work, an LMR planar microfluidic system is used to perform the label-free detection of different concentrations of anti-gliadin antibodies, one of the biomarkers of celiac disease, achieving a limit of detection of 1 μg/mL. The specificity of the sensor to the target analyte is also proved, and the validation of the biofunctionalization process is complemented with an AFM analysis

This publication has 17 references indexed in Scilit: