Safety and Feasibility of Magnetic Resonance Imaging of the Brain at 1.5 Tesla in Patients with Temporary Transmyocardial Pacing Leads
- 28 October 2017
- journal article
- research article
- Published by Georg Thieme Verlag KG in The Thoracic and Cardiovascular Surgeon
- Vol. 67 (02) , 086-091
Abstract
Background Temporary transmyocardial pacing leads (TTPLs) represent an absolute contraindication to magnetic resonance imaging (MRI). The purpose of this study was to evaluate the safety and feasibility of MRI at 1.5 Tesla (T) using a transmit/receive (T/R) head coil in patients with TTPL. Methods TTPLs (220 cm, Osypka TME, Dr. Osypka GmbH, Rheinfelden, Germany) were implanted in a phantom and exposed to conditions of a 1.5 T brain examination using a T/R head coil. Temperature changes at the lead tip were continuously recorded. A total of 28 patients with TTPL and an urgent indication for a brain MRI underwent MRI at 1.5 T with vital sign monitoring. A T/R head coil was used to minimize radiofrequency exposure of the TTPL. Before and immediately after the MRI scan, TTPL lead impedance, pacing capture threshold (PCT), signal slope, and sensing were measured. Serum troponin I was determined before and after MRI to detect thermal myocardial injury. Results In vitro, the maximum temperature increase from radiofrequency-induced heating of the TTPL tip was < 1°C. In vivo, no complications, such as heating sensations, dizziness, unexpected changes in heart rate or rhythm, or other unusual signs or symptoms were observed. No significant changes in the lead impedance, PCT, signal slope, or sensing were recorded. There were no increases of serum troponin I after the MRI examination. Conclusions MRI of the brain may be performed safely at 1.5 T using a T/R head coil in case of an urgent clinical need in patients with TTPL and may be considered a feasible and safe procedure when appropriate precautionary measures are taken.Keywords
This publication has 22 references indexed in Scilit:
- Stroke after Cardiac Surgery and its Association with Asymptomatic Carotid Disease: An Updated Systematic Review and Meta-analysisEuropean Journal of Vascular and Endovascular Surgery, 2011
- Compatibility of temporary pacemaker myocardial pacing leads with magnetic resonance imaging: an ex vivo tissue studyThe International Journal of Cardiovascular Imaging, 2011
- Heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2005
- Induction Ovens and Electromagnetic Interference: What Is the Risk for Patients with Implantable Cardioverter Defibrillators?Journal of Cardiovascular Electrophysiology, 2005
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Markers of Myocardial Damage, Tissue Healing, and Inflammation After Radiofrequency Catheter Ablation of Atrial TachyarrhythmiasJournal of Cardiovascular Electrophysiology, 2004
- Radiofrequency power deposition utilizing thermal imagingMagnetic Resonance in Medicine, 2004
- Safety of MR imaging in patients who have retained metallic materials after cardiac surgery.American Journal of Roentgenology, 1997
- Permanent pacing following repeat cardiac valve surgeryThe American Journal of Cardiology, 1994
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985