Development of a thin high-frequency and high-precision magnetic probe array in Sino-United Spherical Tokamak
- 1 May 2021
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 92 (5) , 053518
Abstract
Since the major/minor radius of the Sino-United Spherical Tokamak (SUNIST) is 0.3/0.23 m, respectively, the space left for magnetic diagnostics in the high field side (HFS) is quite limited. At the same time, a good signal-to-noise ratio and a high-frequency response (up to 1 MHz) are required for equilibrium reconstruction (ER) and Alfven eigenmode studies. Such a magnetic probe array must be extremely thin and tightly close to the central column, not exceeding the inner limiter and leaving the aspect ratio of the spherical tokamak unchanged. Therefore, the turn number and the shape of windings should be highly optimized to enable both a high-frequency response and an enough effective area. A 32-channel magnetic probe array fulfilling these requirements on the central column is designed, calibrated, and installed in the SUNIST. The array consists of 16 probes. Each of them consists of two perpendicular windings, which can measure toroidal and poloidal magnetic fields simultaneously. The effective area and frequency response of each probe are calibrated using a Helmholtz coil and an LCR bridge based on an equivalent probe-and-cable circuit model. After that, an expression of the magnetic diagnostic response to the field coil currents is used to calibrate the installation error. With the full coverage of magnetic probes in the poloidal direction, more reliable ER can be obtained, and the features of magnetohydrodynamic activities in the HFS can be studied.Keywords
Funding Information
- National Natural Science Foundation of China (11827810, 11875177)
- International Atomic Energy Agency (22733)
- Science Challenge Project (TZ2016005)
This publication has 11 references indexed in Scilit:
- Note: Development of a multichannel magnetic probe array for magnetohydrodynamic activity studies in Sino-United Spherical TokamakReview of Scientific Instruments, 2018
- Determination of eddy current response with magnetic measurementsReview of Scientific Instruments, 2017
- Design and calibration of high-frequency magnetic probes for the SUNIST spherical tokamakReview of Scientific Instruments, 2014
- Design of Magnetic Measurement System on SUNIST Spherical TokamakPlasma Science and Technology, 2008
- MHD Equilibrium Reconstruction in the DIII-D TokamakFusion Science and Technology, 2005
- A Research Program of Spherical Tokamak in ChinaPlasma Science and Technology, 2002
- Fast magnetic fluctuation diagnostics for Alfvén eigenmode and magnetohydrodynamics studies at the Joint European TorusReview of Scientific Instruments, 2000
- Phenomenology of major and minor disruptions in high β deuterium and tritium tokamak fusion test reactor plasmaPhysics of Plasmas, 1998
- Reconstruction of current profile parameters and plasma shapes in tokamaksNuclear Fusion, 1985
- Magnetic probes of high frequency responseJournal of Scientific Instruments, 1960