Radiation dose to breast during digital mammography in Tanzania
- 9 January 2024
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 200 (4) , 409-416
Abstract
The aim of this study was to evaluate the mean glandular dose (MGD), to assess the potential for optimization, and to propose diagnostic reference levels (DRLs). MGD was estimated from air kerma measurements and patient information collected during mammography examinations. The 75th percentile values were determined as the third quartile of the median MGD values for all hospitals, and DRLs set as 75th percentile of MGD values. The estimated median values of MGD ranged from 1.5 to 3.9 mGy for craniocaudal projection for median range of 15–59 mm compressed breast thickness (CBT). For a CBT range of 15–63 mm, the median MGD value was 1.5–5.1 mGy for medio-lateral oblique projection. Comparison with other studies showed that the MGD values obtained in this study were relatively high. The magnitude and wide variation of the exposure parameters suggest existing potential for optimization. The training of radiology staff was identified as a top priority.Keywords
Funding Information
- Tanzania Atomic Energy Commission (RP 059)
This publication has 13 references indexed in Scilit:
- AVERAGE GLANDULAR DOSES AND NATIONAL DIAGNOSTIC REFERENCE LEVELS IN MAMMOGRAPHY EXAMINATIONS IN TURKEYRadiation Protection Dosimetry, 2020
- LOW BMI PATIENT DOSE IN DIGITAL RADIOGRAPHYRadiation Protection Dosimetry, 2020
- CHALLENGES IN CLINICAL PRACTICE OF CR MAMMOGRAPHY IN TANZANIARadiation Protection Dosimetry, 2018
- THE CONSISTENCY OF EXPOSURE INDICATOR VALUES IN DIGITAL RADIOGRAPHY SYSTEMSRadiation Protection Dosimetry, 2018
- Optimization of Image Quality and Dose in Digital MammographyJournal of Digital Imaging, 2016
- Diagnostic reference levels for digital mammography in New South WalesJournal of Medical Imaging and Radiation Oncology, 2016
- Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projectsEuropean Journal of Radiology, 2012
- Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detectorBritish Journal of Radiology, 2010
- Solid-state, flat-panel, digital radiography detectors and their physical imaging characteristicsClinical Radiology, 2008
- Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocolPhysics in Medicine & Biology, 2000