Multi-Institutional Verification of a Novel Predictor (Volume-Scaled SUVmax) for Successful Biology-Guided Radiotherapy Delivery of Small Targets
Authors:
- M. Ramish AshrafMAM. Ramish Ashraf
- Daniel PhamDPDaniel Pham
- Girish BalGBGirish Bal
- Huixiao ChenHCHuixiao Chen
- Henry S. ParkHenry S. Park
- Tyler WatkinsTWTyler Watkins
- Bin CaiBCBin Cai
- Shahed N. BadiyanSBShahed N. Badiyan
- Lucas K. VitzthumLVLucas K. Vitzthum
- Billy W. LooBLBilly W. Loo
- Murat SurucuMSMurat Surucu
Open Access
Abstract
Background/Objectives: The aim of this study was to establish the relationship between target size and the required diagnostic PET maximum standard uptake value (SUVmax) thresholds needed for successful Biology-guided Radiotherapy (BgRT) delivery on RefleXion X1 PET-linac. The current clinical eligibility recommendation is an SUVmax ≥ 6 at simulation, but the RefleXion system subsequently evaluates Activity Concentration (AC), which must exceed 5 kBq/mL for successful BgRT planning. Methods: A custom 3D-printed phantom containing six spherical targets (8 to 20 mm diameter) was used with varying target-to-background ratios (5:1 to 20:1) of 18F-FDG to systematically achieve a range of SUVmax values for each target size. Images were acquired on Siemens Biograph mCT for SUVmax quantification and RefleXion X1 for AC measurements. Twenty-four BgRT plans were evaluated, and delivery accuracy was validated using ArcCHECK. Additionally, retrospective data from 18 patients across four institutions were analyzed to validate the phantom-derived findings. Results: The PET-linac successfully planned treatments for 13/24 experiments, all achieving an AC > 5 kBq/mL. SUVmax requirements varied by target size: 16–20 mm targets required an SUVmax > 6, consistent with current recommendations, while smaller targets required higher thresholds (e.g., 13 mm: SUVmax > 10, and 11 mm: SUVmax > 15). 8 and 9 mm targets failed to meet AC requirements even at SUVmax 14. Successful deliveries maintained acceptable accuracy, with gamma passing rates of 92.4% ± 5.0% (3%/2 mm) and 97.6% ± 1.9% (3%/3 mm). Analysis revealed that Volume (cc) × SUVmax > 11 consistently predicted successful BgRT planning across all target sizes. This threshold was validated using multi-institutional PET-CT patient data (mean: 11.36, 95% CI: 9.1–12.9), correctly predicting treatment eligibility in 15 of 18 cases. Conclusions: Target size significantly influences BgRT eligibility. We derived a new criterion, Volume(cc) × SUVmax > 11 (95% CI: 9.1–12).Keywords
This publication has 18 references indexed in Scilit:
- Commissioning of a novel PET‐Linac for biology‐guided radiotherapy (BgRT)Medical Physics, 2024
- BIOGUIDE-X: A First-in-Human Study of the Performance of Positron Emission Tomography-Guided Radiation TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2023
- First-Year Experience of Stereotactic Body Radiation Therapy/Intensity Modulated Radiation Therapy Treatment Using a Novel Biology-Guided Radiation Therapy MachineAdvances in Radiation Oncology, 2023
- Treatment planning system commissioning of the first clinical biology‐guided radiotherapy machineJournal of Applied Clinical Medical Physics, 2022
- A detailed process map for clinical workflow of a new biology‐guided radiotherapy (BgRT) machineJournal of Applied Clinical Medical Physics, 2022
- Beam commissioning of the first clinical biology‐guided radiotherapy systemJournal of Applied Clinical Medical Physics, 2022
- Small‐field measurement and Monte Carlo model validation of a novel image‐guided radiotherapy systemMedical Physics, 2021
- The technical design and concept of a PET/CT linac for biology-guided radiotherapyClinical and Translational Radiation Oncology, 2021
- Total lesion glycolysis: A possible new prognostic parameter in oral cavity squamous cell carcinomaOral Oncology, 2013
- What Is the Best Way to Contour Lung Tumors on PET Scans? Multiobserver Validation of a Gradient-Based Method Using a NSCLC Digital PET PhantomInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012