Efficiency of antiscatter grids for flat-detector CT
Authors:
- Yiannis KyriakouYKYiannis Kyriakou
- Willi KalenderWKWilli Kalender
- 2 October 2007
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 52 (20) , 6275-6293
Abstract
Flat-panel detector CT (FD-CT) scanners offer large volume coverage, but as a consequence are more susceptible to scatter artifacts than standard clinical CT scanners with smaller cone angles. FD-CT scanners can employ antiscatter grids as a scatter rejection technique. We evaluated three standard fluoroscopic antiscatter grids for two different field sizes with respect to scatter suppression efficiency and image quality improvement. The evaluations included simulations and measurements. Regarding the simulation a hybrid model combining deterministic and Monte Carlo (MC) calculations was used combined with an analytical calculation of grid transmission. The scatter-to-primary ratio (SPR) was measured using an adapted collimator technique in order to validate our simulations. The SPR obtained by simulations and measurements with and without antiscatter grids were in agreement typically within 10%. The employment of a grid does not generally provide a significant improvement of the signal-to-noise ratio (SNR). Antiscatter grids led to a significant reduction of cupping artifacts in all cases. There is a trade-off between the SNR and the reduction of the scatter intensity described by the signal-to-noise improvement factor (SNR(if)). For low- or medium-scatter conditions the increase in noise caused by the reduced primary transmission through the grid has to be compensated by a higher exposure. For high scatter conditions SNR(if) is significantly greater than 1; i.e. a decrease of dose of up to 50% can be reached.Keywords
This publication has 21 references indexed in Scilit:
- Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology2004
- Grid and Slot Scan Scatter Reduction in Mammography: Comparison by Using Monte Carlo TechniquesRadiology, 2002
- Performance of glass fiber antiscatter devices at mammographic energiesMedical Physics, 1994
- Contrast and scatter in x-ray imaging.RadioGraphics, 1991
- Studies of performance of antiscatter grids in digital radiography: Effect on signal‐to‐noise ratioMedical Physics, 1990
- Performance of antiscatter grids in diagnostic radiology: Experimental measurements and Monte Carlo simulation studiesMedical Physics, 1985
- A detailed experimental and theoretical comparison of the angular and energy dependencies of grid transmissionMedical Physics, 1983
- Scattered radiation in fan beam imaging systemsMedical Physics, 1982
- Calculation of X-ray grid characteristics by Monte Carlo methodsPhysics in Medicine & Biology, 1982
- Monte Carlo calculations of X-ray scatter data for diagnostic radiologyPhysics in Medicine & Biology, 1981