An Arterial Compliance Sensor for Cuffless Blood Pressure Estimation Based on Piezoelectric and Optical Signals
Open Access
- 16 August 2022
- journal article
- research article
- Published by MDPI AG in Micromachines
- Vol. 13 (8) , 1327
Abstract
Objective: Blood pressure (BP) data can influence therapeutic decisions for some patients, while non-invasive devices that continuously monitor BP can provide patients with a more comprehensive BP assessment. Therefore, this study proposes a multi-sensor-based small cuffless BP monitoring device that integrates a piezoelectric sensor array and an optical sensor, which can monitor the patient’s physiological signals from the radial artery. Method: Based on the Moens–Korteweg (MK) equation of the hemodynamic model, pulse wave velocity (PWV) can be correlated with arterial compliance and BP can be estimated. Therefore, the novel method proposed in this study involves using a piezoelectric sensor array to measure the PWV and an optical sensor to measure the photoplethysmography (PPG) intensity ratio (PIR) signal to estimate the participant’s arterial parameters. The parameters measured by multiple sensors were combined to estimate BP based on the P–β model derived from the MK equation. Result: We recruited 20 participants for the BP monitoring experiment to compare the performance of the BP estimation method with the regression model and the P–β model method with arterial compliance. We then compared the estimated BP with a reference device for validation. The results are presented as the error mean ± standard deviation (SD). Based on the regression model method, systolic blood pressure (SBP) was 0.32 ± 5.94, diastolic blood pressure (DBP) was 2.17 ± 6.22, and mean arterial pressure (MAP) was 1.55 ± 5.83. The results of the P–β model method were as follows: SBP was 0.75 ± 3.9, DBP was 1.1 ± 3.12, and MAP was 0.49 ± 2.82. Conclusion: According to the results of our proposed small cuffless BP monitoring device, both methods of estimating BP conform to ANSI/AAMI/ISO 81060-2:20181_5.2.4.1.2 criterion 1 and 2, and using arterial parameters to calibrate the MK equation model can improve BP estimate accuracy. In the future, our proposed device can provide patients with a convenient and comfortable BP monitoring solution. Since the device is small, it can be used in a public place without attracting other people’s attention, thereby effectively improving the patient’s right to privacy, and increasing their willingness to use it.Keywords
This publication has 57 references indexed in Scilit:
- Association of renal resistive index with aortic pulse wave velocity in hypertensive patientsEuropean Journal of Preventive Cardiology, 2014
- Comparison of sequentially measured Aloka echo-tracking one-point pulse wave velocity with SphygmoCor carotid–femoral pulse wave velocitySage Open Medicine, 2013
- Beer-lambert law along non-linear mean light pathways for the rational analysis of PhotoplethysmographyJournal of Physics: Conference Series, 2010
- Continuous and Noninvasive Blood Pressure Measurement: A Novel Modeling Methodology of the Relationship Between Blood Pressure and Pulse Wave VelocityAnnals of Biomedical Engineering, 2009
- Sensitivity enhancement for colorimetric glucose assays on whole blood by on-chip beam-guidanceBiomedical Microdevices, 2006
- Pulse Transit Time Based on Piezoelectric Technique at the Radial ArteryJournal of Clinical Monitoring and Computing, 2006
- Task Force II: Blood pressure measurement and cardiovacular outcomeBlood Pressure Monitoring, 2001
- Can mean arterial pressure be estimated from measurements of systolic and diastolic blood pressure, and vice versa?Journal Of Hypertension, 1990
- FIR filter design over a discrete powers-of-two coefficient spaceIEEE Transactions on Acoustics, Speech, and Signal Processing, 1983
- Measurements of Young's Modulus of Elasticity of the Canine Aorta with UltrasoundUltrasonic Imaging, 1979