Ultrasound-based transient elastography compared to magnetic resonance elastography in soft tissue-mimicking gels
- 4 November 2009
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (22) , 6979-6990
Abstract
Ultrasound-based transient elastography (TE) and magnetic resonance elastography (MRE) are increasingly used methods for the clinical evaluation of soft tissue mechanical properties and their alteration under diseased conditions. This study proposes a comparison between magnetic resonance elastography (MRE) and ultrasound-based transient elastography (TE). Both methods were tested on the same soft tissue-mimicking gels in a common frequency range in order to allow for direct quantitative comparison. For the four gels tested, relatively good agreement was found between the shear moduli measured by both methods, with an averaged relative difference of 23%. This study demonstrates that under the assumption of homogeneous media that are significantly more elastic than viscous, quantitative results obtained by both methods are comparable.Keywords
This publication has 22 references indexed in Scilit:
- Measurement of liver stiffness with two imaging techniques: Magnetic resonance elastography and ultrasound elastometryJournal of Magnetic Resonance Imaging, 2008
- Magnetic Resonance Elastography for the Noninvasive Staging of Liver FibrosisGastroenterology, 2008
- Non-invasive evaluation of liver fibrosis using transient elastographyJournal of Hepatology, 2008
- Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver diseaseGut, 2007
- In Vivo Determination of Hepatic Stiffness Using Steady-State Free Precession Magnetic Resonance ElastographyInvestigative Radiology, 2006
- Liver fibrosis: non‐invasive assessment with MR elastographyNMR in Biomedicine, 2006
- Diagnosis of cirrhosis by transient elastography (FibroScan): a prospective studyGut, 2006
- Prospective comparison of transient elastography, Fibrotest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis CGastroenterology, 2005
- Diffraction field of a low frequency vibrator in soft tissues using transient elastographyIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1999
- Second-Order Elastic Deformation of SolidsPhysical Review B, 1953