Pressure-Waveform Guidance for Pulmonary Vein Occlusion During Cryoballoon Ablation: Procedural Efficiency and Repeat-Procedure Findings from a Large Single-Center Cohort
Authors:
- Obaida AlothmanObaida Alothman
- Ahmad KeelaniAKAhmad Keelani
- Georgi BorisovGBGeorgi Borisov
- Markus FrommholdMFMarkus Frommhold
- Hytham AbdelwahabHAHytham Abdelwahab
- Gabriele D’AmbrosioGDGabriele D’Ambrosio
- Karam IssaKIKaram Issa
- Sultan AlshehriSASultan Alshehri
- Santi RaffaSRSanti Raffa
- J. Christoph GellerJGJ. Christoph Geller
Open Access
- 25 September 2026
- journal article
- research article
- Published by MDPI AG in Journal of Clinical Medicine
- Vol. 15 (19) , 7462
Abstract
Background/Objectives: The cryoballoon (CB) has simplified pulmonary vein isolation (PVI) for treating atrial fibrillation (AF). Vein occlusion is usually assessed utilizing contrast media which entails possible risks and increases fluoroscopy exposure. Pressure-guided assessment of occlusion has emerged as a potential alternative. The objective of this study is to compare procedural characteristics and repeat-procedure findings of pressure-guided and contrast-guided assessment of vein occlusion. Methods: From July 2018 to December 2023, 1166 patients (485 [41%] female, mean age 65 ± 10 years) with AF (64% paroxysmal) undergoing CB-PVI were enrolled at a single-center, 368 (32%) underwent a contrast-guided and 798 (68%) a pressure-guided approach. Results: Procedure duration (90 ± 20 vs. 93 ± 23 min) and ablation time (1342 ± 361 vs. 1368 ± 406 s) were similar in pressure- and contrast- guided groups, and the acute success rate of PVI was high in both groups. The pressure-guided approach had shorter X-ray times (5.7 ± 3.4 vs. 9.5 ± 4.8 min, p < 0.0001) and lower X-ray doses (82 ± 78.9 vs. 586 ± 546 cGy/cm2, p < 0.0001). In 112 redo-procedures for arrhythmia recurrence after CB-PVI, durable isolation of all veins was observed more often after pressure-guided procedures (28/56 [50%] vs. 13/56 [23%]); per-vein durability was also higher (174/224 [78%] vs. 150/224 [67%], p = 0.015). Conclusions: Pressure-waveform guidance during CB-PVI is a feasible alternative to contrast-guided occlusion assessment and was associated with substantially lower intraprocedural fluoroscopy exposure. Repeat-procedure findings suggest a possible durability signal, but this observation should be considered exploratory and confirmed in prospective studies.Keywords
This publication has 28 references indexed in Scilit:
- Comparison between contrast-guided and pressure-guided ablation using the novel pressure visualization tool for cryoballoon pulmonary vein isolationHeart and Vessels, 2025
- Pressure waveform analysis for occlusion assessment significantly reduces contrast medium use in cryoballoon pulmonary vein isolationJournal of Interventional Cardiac Electrophysiology, 2024
- Meta‐analysis of ultrasound‐guided vs conventional vascular access for cardiac electrophysiology proceduresJournal of Arrhythmia, 2019
- Left atrial voltage mapping: defining and targeting the atrial fibrillation substrateJournal of Interventional Cardiac Electrophysiology, 2019
- Cryoballoon Best Practices II: Practical guide to procedural monitoring and dosing during atrial fibrillation ablation from the perspective of experienced usersHeart Rhythm, 2018
- Individualized cryoballoon energy pulmonary vein isolation guided by real-time pulmonary vein recordings, the randomized ICE-T trialHeart Rhythm, 2017
- Best practice guide for cryoballoon ablation in atrial fibrillation: The compilation experience of more than 3000 proceduresHeart Rhythm, 2015
- Recordings of diaphragmatic electromyograms during cryoballoon ablation for atrial fibrillation accurately predict phrenic nerve injuryHeart Rhythm, 2014
- Phrenic nerve monitoring with diaphragmatic electromyography during cryoballoon ablation for atrial fibrillation: The first human applicationHeart Rhythm, 2011
- Diaphragmatic electromyography during cryoballoon ablation: a novel concept in the prevention of phrenic nerve palsyHeart Rhythm, 2011