A simulation study on proton computed tomography (CT) stopping power accuracy using dual energy CT scans as benchmark
- 29 July 2015
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Oncologica
- Vol. 54 (9) , 1638-1642
Abstract
Background. Accurate stopping power estimation is crucial for treatment planning in proton therapy, and the uncertainties in stopping power are currently the largest contributor to the employed dose margins. Dual energy x-ray computed tomography (CT) (clinically available) and proton CT (in development) have both been proposed as methods for obtaining patient stopping power maps. The purpose of this work was to assess the accuracy of proton CT using dual energy CT scans of phantoms to establish reference accuracy levels. Material and methods. A CT calibration phantom and an abdomen cross section phantom containing inserts were scanned with dual energy and single energy CT with a state-of-the-art dual energy CT scanner. Proton CT scans were simulated using Monte Carlo methods. The simulations followed the setup used in current prototype proton CT scanners and included realistic modeling of detectors and the corresponding noise characteristics. Stopping power maps were calculated for all three scans, and compared with the ground truth stopping power from the phantoms. Results. Proton CT gave slightly better stopping power estimates than the dual energy CT method, with root mean square errors of 0.2% and 0.5% (for each phantom) compared to 0.5% and 0.9%. Single energy CT root mean square errors were 2.7% and 1.6%. Maximal errors for proton, dual energy and single energy CT were 0.51%, 1.7% and 7.4%, respectively. Conclusion. Better stopping power estimates could significantly reduce the range errors in proton therapy, but requires a large improvement in current methods which may be achievable with proton CT.Keywords
This publication has 25 references indexed in Scilit:
- A method for selection of beam angles robust to intra-fractional motion in proton therapy of lung cancerActa Oncologica, 2014
- Selection of patients for radiotherapy with protons aiming at reduction of side effects: The model-based approachRadiotherapy and Oncology, 2013
- The model-based approach to clinical studies in particle radiotherapy – A new concept in evidence based radiation oncology?Radiotherapy and Oncology, 2013
- Implementation and initial clinical experience of offline PET/CT-based verification of scanned carbon ion treatmentRadiotherapy and Oncology, 2013
- Feasibility study of in vivo MRI based dosimetric verification of proton end-of-range for liver cancer patientsRadiotherapy and Oncology, 2013
- Robustness of IMPT treatment plans with respect to inter-fractional set-up uncertainties: Impact of various beam arrangements for cranial targetsActa Oncologica, 2012
- Range uncertainties in proton therapy and the role of Monte Carlo simulationsPhysics in Medicine & Biology, 2012
- The calibration of CT Hounsfield units for radiotherapy treatment planningPhysics in Medicine & Biology, 1996
- Computed tomography using proton energy lossPhysics in Medicine & Biology, 1981
- Quantitative proton tomography: preliminary experimentsPhysics in Medicine & Biology, 1976