Comparison of surface dose during whole breast radiation therapy on Halcyon and TrueBeam using Cherenkov imaging
- 6 March 2023
- conference paper
- Published by SPIE-Intl Soc Optical Eng in Proceedings of SPIE
- Vol. 12371, 23-34
Abstract
The emergence of the Halcyon linear accelerator has allowed for increased patient throughput and improved treatment times for common treatment sites in radiation oncology. However, it has been shown that this can lead to increased surface dose in sites like breast cancer compared with treatments on conventional machines with flattened radiation beams. Cherenkov imaging can be used to estimate surface dose by detection of Cherenkov photons emitted in proportion to energy deposition from high energy electrons in tissue. Phantom studies were performed with both square beams in reference conditions and with clinical treatments, and dosimeter readings and Cherenkov images report higher surface dose (25% for flat phantom entrance dose, 5.9% for breast phantom treatment) from Halcyon beam deliveries than for equivalent deliveries from a TrueBeam linac. Additionally, the first Cherenkov images of a patient treated with Halcyon were acquired, and superficial dose was estimated.Keywords
This publication has 18 references indexed in Scilit:
- Signal intensity analysis and optimization for in vivo imaging of Cherenkov and excited luminescencePhysics in Medicine & Biology, 2018
- Evaluation of the dosimetric characteristics of 6 MV flattened and unflattened photon beamJournal of King Saud University – Science, 2017
- Beam and tissue factors affecting Cherenkov image intensity for quantitative entrance and exit dosimetry on human tissueJournal of Biophotonics, 2016
- Flattening filter‐free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work GroupJournal of Applied Clinical Medical Physics, 2015
- Validation of OSLD and a treatment planning system for surface dose determination in IMRT treatmentsMedical Physics, 2014
- Cherenkov Video Imaging Allows for the First Visualization of Radiation Therapy in Real TimeInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2014
- Optical dosimetry of radiotherapy beams using Cherenkov radiation: the relationship between light emission and dosePhysics in Medicine & Biology, 2014
- Superficial dosimetry imaging based on Čerenkov emission for external beam radiotherapy with megavoltage x‐ray beamMedical Physics, 2013
- Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA‐compliant phase space calculationsMedical Physics, 2011
- Dosimetric characteristics of unflattened 6MV photon beams of a clinical linear accelerator: A Monte Carlo studyApplied Radiation and Isotopes, 2007