Absorbable Plate Strength Loss During Molding
Authors:
- Kevin J. KellyKKKevin J. Kelly
- Victor ZaydfudimVZVictor Zaydfudim
- Noah L. WalcuttNWNoah L. Walcutt
- Soheil S. LahijaniSLSoheil S. Lahijani
- R. Bruce ShackRSR. Bruce Shack
- Wesley P. ThayerWTWesley P. Thayer
- 1 May 2010
- journal article
- Published by Wolters Kluwer Health in Journal of Craniofacial Surgery
- Vol. 21 (3) , 644-647
Abstract
Bioabsorbable plating systems play an integral role in cranial vault remodeling. After experiencing a case of plate failure requiring emergent reexploration, we investigated the potential causes. We hypothesize that extended submersion in the molding bath during plate preparation might advance the rate of hydrolysis and compromise plate structural integrity. Using an absorbable poly-D/L-lactic acid plating system, we assessed the effect of extended submersion on plate strength and stiffness when loaded in a cantilever fashion and with pure tension. We assessed these differences with the Student t test and linear regression modeling. We also generated a computer model of the plates for finite element analysis. When left in the molding bath for extended periods, the plates changed color and lost strength. After 5 minutes, 30% of maximum plate load capacity was lost in a cantilever beam test (P < 0.001) consistent with use of a 15% thinner plate. Tensile testing revealed the initial elastic modulus of 6.42 +/- 0.13 GPa decreased 16% to 5.41 +/- 0.50 GPa after 5 minutes of submersion (P = 0.027). The changes in plate strength and elastic modulus both worsened with increased submersion times. Finite element analysis of the plates also predicted clinically significant increases in plate deviation under normal loading conditions. Our study demonstrates that extended submersion of absorbable plates during molding results in a significant loss of plate strength and stiffness. Further, our computer model predicts that these changes could result in an unacceptable plate deviation under normal loading conditions. Together, these data caution against overmolding of plates to avoid compromising their structural integrity.Keywords
This publication has 10 references indexed in Scilit:
- Maxillary and Mandibular Osteosyntheses with PLGA and P(L/DL)LA Implants: A 5-Year Inpatient Biocompatibility and Degradation ExperiencePlastic and Reconstructive Surgery, 2006
- Resorbable PLLA-PGA Plate and Screw Fixation in Pediatric Craniofacial Surgery: Clinical Experience in 1883 PatientsPlastic and Reconstructive Surgery, 2004
- Early Failure of Absorbable Plating in a Patient With Syndromic BrachycephalyJournal of Craniofacial Surgery, 2004
- Outcome Analysis for Correction of Single Suture Craniosynostosis Using Resorbable FixationJournal of Craniofacial Surgery, 2001
- Osteotomy site healing following mandibular sagittal split osteotomy and rigid fixation with polylactide biodegradable screwsInternational Journal of Oral and Maxillofacial Surgery, 1999
- A Longitudinal, Statistical Study of Reoperation Rates in CraniosynostosisPlastic and Reconstructive Surgery, 1997
- BioabsorbablesJournal of Craniofacial Surgery, 1997
- Bioabsorbable Plates and Screws in Pediatric Craniofacial SurgeryJournal of Craniofacial Surgery, 1997
- Absorbable, Invisible, and Flexible Plating System for the Craniofacial SkeletonJournal of Craniofacial Surgery, 1997
- Foreign body reactions to resorbable poly(l-lactide) bone plates and screws used for the fixation of unstable zygomatic fracturesJournal of Oral and Maxillofacial Surgery, 1993