Abstract
This paper presents the design, fabrication, calibration, and comprehensive characterization of a homemade tri-axial fluxgate magnetometer. The magnetometer, utilizing a ring core configuration, was developed to measure ultra-low magnetic fields with high sensitivity and stability. Critical stages from material selection to sensor geometry optimization are discussed in detail. A series of critical characterization processes were conducted, including zero-field voltage determination, scale factor calculation, resolution measurement, noise analysis, bias assessment, cross-field effect evaluation, temperature dependency, and bandwidth determination. The sensor demonstrated a minimum detectable magnetic field resolution of 2.2 nT with a noise level of 1.1 nT/√Hz at 1 Hz. Temperature dependency tests revealed minimal impact on sensor output with a maximum shift of 120 nT in the range of 60 °C, which was effectively compensated through calibration to less than 5 nT. Additionally, the paper introduces a model function in matrix form to relate the magnetometer’s output voltage to the measured magnetic field, incorporating temperature dependency and cross-field effects. This work highlights the importance of meticulous calibration and optimization in developing fluxgate magnetometers suitable for various applications, from space exploration to biomedical diagnostics.
Funding Information
  • TUBITAK (The Scientific and Technological Research Council of Turkey) (218M647, 123N773)
  • Slovak Academy of Sciences (218M647, 123N773)
  • 1003 programme of TÜBİTAK (218M647, 123N773)
  • Slovak Academy of Sciences and TÜBİTAK (218M647, 123N773)

This publication has 17 references indexed in Scilit: