Initial experience with the TightRail™ Rotating Mechanical Dilator Sheath for transvenous lead extraction
- 14 October 2015
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 18 (7) , 1043-1048
Abstract
Aims In parallel with increasing implantation rates and patients' longer life expectancy, the need for transvenous lead extraction (TLE) as a specialized procedure has shown a significant growth over years. Herein, we aimed to present our initial experience in TLE by using a novel TightRail™ Rotating Mechanical Dilator Sheath. Methods and results Between October 2014 and March 2015, a total of 42 leads in 23 patients were removed at our tertiary referral centre. All of the extracted leads were >12 months old and indications for extraction were based on the recommendations of the Heart Rhythm Society. The leads were removed by using the TightRail™ Mechanical Dilator Sheath (Spectranetics Corporation) with the rotational cutting force only. Indications for lead removal included cardiac device infection in 12 (52.2%) cases, lead malfunction in the 10 (43.5%) cases, and upgrade to cardiac resynchronization therapy-defibrillator (CRT-D) in the remaining 1 case (4.3%). The extracted devices were pacemaker in 10 (43.4%) cases, implantable cardioverter-defibrillator (ICD) in 7 (30.4%) cases, and CRT in the remaining 6 (26.0%) subjects. Among 42 leads, 10 (23.8%) were right ventricular, 14 (33.3%) were atrial, 13 (31.0%) were defibrilator, and 5 (11.9%) were coronary sinus electrodes. The median time from implantation was 72 (18–216) months. Complete procedural success with TightRail™ system alone was achieved in 22 (95.7%) patients (41/42 leads) and overall clinical success was 100%. One right ventricular lead was completely removed with the help of femoral snare. All the patients were discharged uneventfully without any complication. Conclusion Our preliminary data with small sample size show that TightRail™ Mechanical Dilator Sheath is a new useful tool for chronically implanted pacemaker (PM)/ICD leads. Continued investigation including large patient cohort is required to evaluate success and complication rates in comparison to other tools and techniques.Keywords
This publication has 18 references indexed in Scilit:
- Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009JACC, 2012
- Long-Term Mortality After Transvenous Lead ExtractionCirculation: Arrhythmia and Electrophysiology, 2012
- Implantable cardioverter-defibrillator utilizationEP Europace, 2011
- The challenges of transvenous lead extractionHeart, 2011
- Percutaneous extraction of cardiac pacemaker and implantable cardioverter defibrillator leads with evolution mechanical dilator sheath: a single-centre experienceEP Europace, 2010
- Update on Cardiovascular Implantable Electronic Device Infections and Their ManagementCirculation, 2010
- Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient ManagementHeart Rhythm, 2009
- Extraction of Pacing Leads: Overview of Current TechniquesJournal of Cardiovascular Electrophysiology, 2006
- Diagnosis and Management of Infections Involving Implantable Electrophysiologic Cardiac DevicesAnnals of Internal Medicine, 2000
- Five‐Years Experience with Intravascular Lead ExtractionPacing and Clinical Electrophysiology, 1994