Patient-tailored implantable cardioverter defibrillator testing using the upper limit of vulnerability: the TULIP protocol
Authors:
- Bernd LemkeBLBernd Lemke
- on behalf of the TULIP (Threshold test using Upper Limit during ImPlantation) Study GroupOGon behalf of the TULIP (Threshold test using Upper Limit during ImPlantation) Study Group
Open Access
- 19 April 2008
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 10 (8) , 907-913
Abstract
We evaluated the feasibility of the TULIP (Threshold test using Upper Limit during ImPlantation) protocol, which was designed to provide a confirmed, low defibrillation energy value during implantable cardioverter defibrillator (ICD) implantation with only two induced ventricular fibrillation (VF) episodes. Ninety-eight patients (62 ± 12 years, 86 male) from 13 clinical centres underwent an active can ICD implantation. A single coupling interval derived from electrocardiogram lead II during ventricular pacing was used for VF induction shocks at 13, 11, 9, and 6 J in a step-down manner until the upper limit of VF induction (ULVI) was determined. If ULVI ≥9 J, a defibrillation energy of ULVI + 4 J was tested. For ULVI <9 J, the defibrillation test energy was 9 J. In 79/98 patients (80.6%), two induced VF episodes were sufficient to obtain confirmed defibrillation energy of 11.1 ± 3.3 J. The mean strength of the successful VF induction shock was 6.8 ± 4.3 J, the coupling interval was 303 ± 35 ms, and the number of delivered induction shocks until the first VF induction was 3.9 ± 1.6. TULIP is a safe and simple device testing procedure allowing the determination of confirmed, low defibrillation energy in most patients with two VF episodes induced at a single coupling interval.Keywords
This publication has 26 references indexed in Scilit:
- Inductionless or Limited Shock Testing Is Possible in Most Patients With Implantable Cardioverter- Defibrillators/Cardiac Resynchronization Therapy DefibrillatorsCirculation, 2007
- The Dilemma of ICD Implant TestingPacing and Clinical Electrophysiology, 2007
- Effect of defibrillation testing on management during implantable cardioverter-defibrillator implantationAmerican Heart Journal, 2006
- Defibrillation efficacy testing: Long-term follow-up and mortalityEP Europace, 2005
- Intraoperative Testing of the Implantable Cardioverter‐Defibrillator: How Much Is Enough?Journal of Cardiovascular Electrophysiology, 2005
- One conversion of ventricular fibrillation is adequate for implantable cardioverter-defibrillator implant: An analysis from the Low Energy Safety Study (LESS)Heart Rhythm, 2005
- Comparison of Step‐Down and Binary Search Algorithms for Determination of Defibrillation Threshold in HumansPacing and Clinical Electrophysiology, 2004
- Prospective randomized comparison of two defibrillation safety margins in unipolar, active pectoral defibrillator therapy.Pacing and Clinical Electrophysiology, 2003
- Efficacy and Temporal Stability of Reduced Safety Margins for Ventricular DefibrillationCirculation, 2002
- Safety and efficacy of implantable defibrillator therapy with programmed shock energy at twice the augmented step-down defibrillation threshold: results of the prospective, randomized, multicenter low-energy endotak trialThe American Journal of Cardiology, 1999