Abstract
In ultrasound imaging catheter applications, gathering the data from multi-element transducer arrays is difficult as there is a restriction on cable count due to the diameter of the catheter. In such applications, CMUT-on-CMOS technology allows for 2D arrays with many elements to be designed and bonded directly onto CMOS circuitry. This allows for complex electronics to be placed at the tip of the catheter which leads to the possibility to include electronic multiplexing techniques to greatly reduce the cable count required for a large element array. Current approaches to cable reduction tend to rely on area and power hungry circuits to function, making them unsuitable for use in catheters. Furthermore the length requirement for catheters and lack of power available to on-chip cable drivers leads to limited signal strength at the receiver end. In this paper an alternative approach using Analogue Time Division Multiplexing (TDM) is presented, which addresses the cable restrictions of the catheter and, using a novel digital demultiplexing technique, allows for a reduction in the number of analogue signal processing stages required.Keywords
This publication has 6 references indexed in Scilit:
- Design of frequency-division multiplexing front-end receiver electronics for CMUT-on-CMOS based intracardiac echocardiography2014
- CMUT-in-CMOS 2D arrays with advanced multiplexing and time-gain control2014
- A volumetric CMUT-based ultrasound imaging system simulator with integrated reception and μ-beamforming electronics modelsIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- Front-end receiver electronics for high-frequency monolithic CMUT-on-CMOS imaging arraysIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2011
- Fully sampled matrix transducer for real time 3D ultrasonic imaging2004
- A 120 MHz, 12 mW CMOS current feedback opamp2002