Toward real‐time verification for MLC tracking treatments using time‐resolved EPID imaging
- 22 December 2020
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (3) , 953-964
Abstract
Purpose: In multileaf collimator (MLC) tracking, the MLC positions from the original treatment plan are continuously modified to account for intrafraction tumor motion. As the treatment is adapted in real time, there is additional risk of delivery errors which cannot be detected using traditional pretreatment dose verification. The purpose of this work is to develop a system for real‐time geometric verification of MLC tracking treatments using an electronic portal imaging device (EPID).Methods: MLC tracking was utilized during volumetric modulated arc therapy (VMAT). During these deliveries, treatment beam images were taken at 9.57 frames per second using an EPID and frame grabber computer. MLC positions were extracted from each image frame and used to assess delivery accuracy using three geometric measures: the location, size, and shape of the radiation field. The EPID‐measured field location was compared to the tumor motion measured by implanted electromagnetic markers. The size and shape of the beam were compared to the size and shape from the original treatment plan, respectively. This technique was validated by simulating errors in phantom test deliveries and by comparison between EPID measurements and treatment log files. The method was applied offline to images acquired during the LIGHT Stereotactic Ablative Body Radiotherapy (SABR) clinical trial, where MLC tracking was performed for 17 lung cancer patients. The EPID‐based verification results were subsequently compared to post‐treatment dose reconstruction.Results: Simulated field location errors were detected during phantom validation tests with an uncertainty of 0.28 mm (parallel to MLC motion) and 0.38 mm (perpendicular), expressed as a root‐mean‐square error (RMSError). For simulated field size errors, the RMSError was 0.47 cm2 and field shape changes were detected for random errors with standard deviation ≥ 2.5 mm. For clinical lung SABR deliveries, field location errors of 1.6 mm (parallel MLC motion) and 4.9 mm (perpendicular) were measured (expressed as a full‐width‐half‐maximum). The mean and standard deviation of the errors in field size and shape were 0.0 ± 0.3 cm2 and 0.3 ± 0.1 (expressed as a translation‐invariant normalized RMS). No correlation was observed between geometric errors during each treatment fraction and dosimetric errors in the reconstructed dose to the target volume for this cohort of patients.Conclusion: A system for real‐time delivery verification has been developed for MLC tracking using time‐resolved EPID imaging. The technique has been tested offline in phantom‐based deliveries and clinical patient deliveries and was used to independently verify the geometric accuracy of the MLC during MLC tracking radiotherapy.Keywords
Funding Information
- Varian Medical Systems
- National Health and Medical Research Council
This publication has 32 references indexed in Scilit:
- An independent system for real-time dynamic multileaf collimation trajectory verification using EPIDPhysics in Medicine & Biology, 2013
- A system for EPID‐based real‐time treatment delivery verification during dynamic IMRT treatmentMedical Physics, 2013
- Model‐based prediction of portal dose images during patient treatmentMedical Physics, 2013
- Image-Based Dynamic Multileaf Collimator Tracking of Moving Targets During Intensity-Modulated Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2012
- Dosimetric Effect of Intrafraction Motion and Residual Setup Error for Hypofractionated Prostate Intensity-Modulated Radiotherapy With Online Cone Beam Computed Tomography Image GuidanceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Using an EPID for patient‐specific VMAT quality assuranceMedical Physics, 2011
- Electromagnetic-Guided Dynamic Multileaf Collimator Tracking Enables Motion Management for Intensity-Modulated Arc TherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2011
- Clinical significance of multi-leaf collimator positional errors for volumetric modulated arc therapyRadiotherapy and Oncology, 2010
- Online prediction of respiratory motion: multidimensional processing with low-dimensional feature learningPhysics in Medicine & Biology, 2010
- Dosimetric Consequences of Intrafraction Prostate MotionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008