Time to take the gloves off: the use of radiation reduction gloves can greatly increase patient dose
Open Access
- 8 November 2014
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 15 (6) , 351-359
Abstract
Sterile radiation reduction gloves have been widely used over the past several decades in an effort to reduce hand doses during fluoroscopically guided procedures. While multiple studies have quantified the potential dose reduction to the fluoroscope operator from the use of such gloves, possible effects on the patient have not yet been quantified. The aim of this study was to examine the impact on patient dose when radiation reduction gloves are used. The impact on patient dose when using radiation reduction gloves in the field of view (FOV) was evaluated by measuring patient entrance surface dose rates (ESDR) using three C‐arm fluoroscopes for a range of patient sizes and different operating and magnification modes. Multiple measuring fields were used in combination with both peripheral and central glove placement. ESDR were measured with no glove in the FOV, with one radiation reduction glove, and with double gloves in the FOV, to replicate the actions of some fluoroscope operators. Compared to an ungloved hand, the use of a single radiation reduction glove in the measuring field resulted in up to a 2.8‐fold increase in ESDR. The use of double radiation reduction gloves resulted in up to a 4.9‐fold increase in ESDR. In both cases, the increase in ESDR was dependent on the size of the patient and on the operating and magnification modes used, and ranged from no increase up to the aforementioned maximum. When used in the FOV, and particularly within the measuring field, radiation reduction gloves can substantially increase ESDR. This increase in dose, when considered against the relatively small published reduction in dose to the operator's hands, may mean that the increased risks from the use of radiation reduction gloves outweigh the benefits. In any case, hands should not be placed in the FOV if not required by the goals of the procedure.PACS number: 87.59.C‐Keywords
This publication has 16 references indexed in Scilit:
- ICRP Publication 121: Radiological Protection in Paediatric Diagnostic and Interventional RadiologyAnnals of the ICRP, 2013
- Functionality and operation of fluoroscopic automatic brightness control/automatic dose rate control logic in modern cardiovascular and interventional angiography systems: A Report of Task Group 125 Radiography/Fluoroscopy Subcommittee, Imaging Physics Committee, Science CouncilMedical Physics, 2012
- Radiological Protection in Fluoroscopically Guided Procedures Performed Outside the Imaging DepartmentAnnals of the ICRP, 2010
- Operator Shielding: How and WhyTechniques in Vascular and Interventional Radiology, 2010
- Occupational Radiation Protection in Interventional Radiology: A Joint Guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional RadiologyJournal of Vascular and Interventional Radiology, 2010
- Fluoroscopically guided percutaneous vertebroplasty: Assessment of radiation doses and implementation of procedural routines to reduce operator exposureActa Radiologica, 2009
- The operation logic of automatic dose control of fluoroscopy system in conjunction with spectral shaping filtersMedical Physics, 2007
- Radiation exposure to the hands of operators during angiographic proceduresEuropean Journal of Radiology, 1995
- Finger doses during interventional radiology: The value of flexible protective glovesRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1991
- Composite Materials for X-ray ProtectionHealth Physics, 1991