Standard MRI-based attenuation correction for PET/MRI phantoms: a novel concept using MRI-visible polymer
Authors:
Open Access
- 18 February 2021
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 8 (1) , 1-13
Abstract
Background: PET/MRI phantom studies are challenged by the need of phantom-specific attenuation templates to account for attenuation properties of the phantom material. We present a PET/MRI phantom built from MRI-visible material for which attenuation correction (AC) can be performed using the standard MRI-based AC. Methods: A water-fillable phantom was 3D-printed with a commercially available MRI-visible polymer. The phantom had a cylindrical shape and the fillable compartment consisted of a homogeneous region and a region containing solid rods of different diameters. The phantom was filled with a solution of water and [18F]FDG. A 30 min PET/MRI acquisition including the standard Dixon-based MR-AC method was performed. In addition, a CT scan of the phantom was acquired on a PET/CT system. From the Dixon in-phase, opposed-phase and fat images, a phantom-specific AC map (Phantom MR-AC) was produced by separating the phantom material from the water compartment using a thresholding-based method and assigning fixed attenuation coefficients to the individual compartments. The PET data was reconstructed using the Phantom MR-AC, the original Dixon MR-AC, and an MR-AC just containing the water compartment (NoWall-AC) to estimate the error of ignoring the phantom walls. CT-based AC was employed as the reference standard. Average %-differences in measured activity between the CT corrected PET and the PET corrected with the other AC methods were calculated. Results: The phantom housing and the liquid compartment were both visible and distinguishable from each other in the Dixon images and allowed the segmentation of a phantom-specific MR-based AC. Compared to the CT-AC PET, average differences in measured activity in the whole water compartment in the phantom of −0.3%, 9.4%, and −24.1% were found for Dixon phantom MR-AC, MR-AC, and NoWall-AC based PET, respectively. Average differences near the phantom wall in the homogeneous region were −0.3%, 6.6%, and −34.3%, respectively. Around the rods, activity differed from the CT-AC PET by 0.7%, 8.9%, and −45.5%, respectively. Conclusion: The presented phantom material is visible using standard MR sequences, and thus, supports the use of standard, phantom-independent MR measurements for MR-AC in PET/MRI phantom studies.Keywords
Funding Information
- H2020 Marie Skłodowska-Curie Actions (No. 764458)
- Austrian Science Fund (I3451-N32)
- Österreichische Forschungsförderungsgesellschaft (858060)
This publication has 40 references indexed in Scilit:
- Magnetic Resonance-Based Attenuation Correction and Scatter Correction in Neurological Positron Emission Tomography/Magnetic Resonance Imaging—Current Status With Emerging ApplicationsFrontiers in Physics, 2020
- Combined PET/MRI: Global Warming—Summary Report of the 6th International Workshop on PET/MRI, March 27–29, 2017, Tübingen, GermanyMolecular Imaging and Biology, 2017
- Technical and instrumentational foundations of PET/MRIEuropean Journal of Radiology, 2017
- A multi-centre evaluation of eleven clinically feasible brain PET/MRI attenuation correction techniques using a large cohort of patientsNeuroImage, 2017
- Advances in PET/MR instrumentation and image reconstructionBritish Journal of Radiology, 2016
- MR Imaging–Guided Attenuation Correction of PET Data in PET/MR ImagingPET Clinics, 2016
- Investigating the state-of-the-art in whole-body MR-based attenuation correction: an intra-individual, inter-system, inventory study on three clinical PET/MR systemsMagnetic Resonance Materials in Physics, Biology and Medicine, 2016
- MR‐based field‐of‐view extension in MR/PET: B0 homogenization using gradient enhancement (HUGE)Magnetic Resonance in Medicine, 2012
- MRI for attenuation correction in PET: methods and challengesMagnetic Resonance Materials in Physics, Biology and Medicine, 2012
- Tissue Classification as a Potential Approach for Attenuation Correction in Whole-Body PET/MRI: Evaluation with PET/CT DataJournal of Nuclear Medicine, 2009