Multileaf collimator characterization and modeling for a 1.5 T MR-linac using static synchronous and asynchronous sweeping gaps
- 14 March 2024
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 69 (7) , 075004
Abstract
Objective. The Elekta unity MR-linac delivers step-and-shoot intensity modulated radiotherapy plans using a multileaf collimator (MLC) based on the Agility MLC used on conventional Elekta linacs. Currently, details of the physical Unity MLC and the computational model within its treatment planning system (TPS) Monaco are lacking in published literature. Recently, a novel approach to characterize the physical properties of MLCs was introduced using dynamic synchronous and asynchronous sweeping gap (aSG) tests. Our objective was to develop a step-and-shoot version of the dynamic aSG test to characterize the Unity MLC and the computational MLC models in the Monaco and RayStation TPSs. Approach. Dynamic aSG were discretized into a step-and-shoot aSG by investigating the number of segments/sweep and the minimal number of monitor units (MU) per segment. The step-and-shoot aSG tests were compared to the dynamic aSG tests on a conventional linac at a source-to-detector distance of 143.5 cm, mimicking the Unity configuration. the step-and-shoot aSG tests were used to characterize the Unity MLC through measurements and dose calculations in both TPSs. Main results. The step-and-shoot aSGs tests with 100 segments and 5 MU/segment gave results very similar to the dynamic aSG experiments. The effective tongue-and-groove width of the Unity gradually increased up to 1.4 cm from the leaf tip end. The MLC models in RayStation and Monaco agreed with experimental data within 2.0% and 10%, respectively. The largest discrepancies in Monaco were found for aSG tests with >10 mm leaf interdigitation, which are non-typical for clinical plans. Significance. The step-and-shoot aSG tests accurately characterize the MLC in step-and-shoot delivery mode. The MLC model in RayStation 2023B accurately describes the tongue-and-groove and leaf tip effects whereas Monaco overestimates the tongue-and-groove shadowing further away from the leaf tip end.Keywords
This publication has 30 references indexed in Scilit:
- Transmission probability filter optimization for Agility MLC in Monaco treatment planning systemJournal of Applied Clinical Medical Physics, 2023
- AAPM MEDICAL PHYSICS PRACTICE GUIDELINE 5.b: Commissioning and QA of treatment planning dose calculations—Megavoltage photon and electron beamsJournal of Applied Clinical Medical Physics, 2022
- Acceptance procedure for the linear accelerator component of the 1.5 T MRI‐linacJournal of Applied Clinical Medical Physics, 2021
- Machine QA for the Elekta Unity system: A Report from the Elekta MR‐linac consortiumMedical Physics, 2021
- Performance of a multileaf collimator system for a 1.5T MR‐linacMedical Physics, 2020
- Commissioning of a 1.5T Elekta Unity MR‐linac: A single institution experienceJournal of Applied Clinical Medical Physics, 2020
- Impact of the MLC leaf‐tip model in a commercial TPS: Dose calculation limitations and IROC‐H phantom failuresJournal of Applied Clinical Medical Physics, 2020
- Beam modeling and beam model commissioning for Monte Carlo dose calculation‐based radiation therapy treatment planning: Report of AAPM Task Group 157Medical Physics, 2019
- Monte Carlo and analytic modeling of an Elekta Infinity linac with Agility MLC: Investigating the significance of accurate model parameters for small radiation fieldsJournal of Applied Clinical Medical Physics, 2018
- Agility MLC transmission optimization in the Monaco treatment planning systemJournal of Applied Clinical Medical Physics, 2018