Parallel transmit pulse design for patients with deep brain stimulation implants
Open Access
- 19 June 2014
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 73 (5) , 1896-1903
Abstract
Purpose Specific absorption rate (SAR) amplification around active implantable medical devices during diagnostic MRI procedures poses a potential risk for patient safety. In this study, we present a parallel transmit (pTx) strategy that can be used to safely scan patients with deep brain stimulation (DBS) implants. Methods We performed electromagnetic simulations at 3T using a uniform phantom and a multitissue realistic head model with a generic DBS implant. Our strategy is based on using implant‐friendly modes, which are defined as the modes of an array that reduce the local SAR around the DBS lead tip. These modes are used in a spokes pulse design algorithm in order to produce highly uniform magnitude least‐squares flip angle excitations. Results Local SAR (1 g) at the lead tip is reduced below 0.1 W/kg compared with 31.2 W/kg, which is obtained by a simple quadrature birdcage excitation without any sort of SAR mitigation. For the multitissue realistic head model, peak 10 g local SAR and global SAR are obtained as 4.52 W/kg and 0.48 W/kg, respectively. A uniform axial flip angle is also obtained (NRMSE <3%). Conclusion Parallel transmit arrays can be used to generate implant‐friendly modes and to reduce SAR around DBS implants while constraining peak local SAR and global SAR and maximizing flip angle homogeneity. Magn Reson Med 73:1896–1903, 2015.Keywords
This publication has 29 references indexed in Scilit:
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- Reduction of implant RF heating through modification of transmit coil electric fieldMagnetic Resonance in Medicine, 2010
- Designing passive MRI‐safe implantable conducting leads with electrodesMedical Physics, 2010
- New method to monitor RF safety in MRI‐guided interventions based on RF induced image artefactsMedical Physics, 2010
- Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurementsMagnetic Resonance in Medicine, 2008
- Permanent Neurological Deficit Related to Magnetic Resonance Imaging in a Patient with Implanted Deep Brain Stimulation Electrodes for Parkinson's Disease: Case ReportNeurosurgery, 2005
- Simple design changes to wires to substantially reduce MRI-induced heating at 1.5 T: implications for implanted leadsMagnetic Resonance Imaging, 2005
- RF safety of wires in interventional MRI: Using a safety indexMagnetic Resonance in Medicine, 2001
- A Green's function approach to local rf heating in interventional MRIMedical Physics, 2001
- On the heating of linear conductive structures as guide wires and catheters in interventional MRIJournal of Magnetic Resonance Imaging, 2001