Investigating the role of imaging factors in the variability of CT‐based texture analysis metrics
Open Access
- 14 November 2023
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (4) , e14192
Abstract
Objective: This study assesses the robustness of first‐order radiomic texture features namely interquartile range (IQR), coefficient of variation (CV) and standard deviation (SD) derived from computed tomography (CT) images by varying dose, reconstruction algorithms and slice thickness using scans of a uniform water phantom, a commercial anthropomorphic liver phantom, and a human liver in‐vivo.Materials and Methods: Scans were acquired on a 16 cm detector GE Revolution Apex Edition CT scanner with variations across three different nominal slice thicknesses: 0.625, 1.25, and 2.5 mm, three different dose levels: CTDIvol of 13.86 mGy for the standard dose, 40% reduced dose and 60% reduced dose and two different reconstruction algorithms: a deep learning image reconstruction (DLIR‐high) algorithm and a hybrid iterative reconstruction (IR) algorithm ASiR‐V50% (AV50) were explored, varying one at a time. To assess the effect of non‐linear modifications of images by AV50 and DLIR‐high, images of the water phantom were also reconstructed using filtered back projection (FBP). Quantitative measures of IQR, CV and SD were extracted from twelve pre‐selected, circular (1 cm diameter) regions of interest (ROIs) capturing different texture patterns across all scans.Results: Across all scans, imaging, and reconstruction settings, CV, IQR and SD were observed to increase with reduction in dose and slice thickness. An exception to this observation was found when using FBP reconstruction. Lower values of CV, IQR and SD were observed in DLIR‐high reconstructions compared to AV50 and FBP. The Poisson statistics were more stringently noted in FBP than DLIR‐high and AV50, due to the non‐linear nature of the latter two algorithms.Conclusion: Variation in image noise due to dose reduction algorithms, tube current, and slice thickness show a consistent trend across phantom and patient scans. Prospective evaluation across multiple centers, scanners and imaging protocols is needed for establishing quality assurance standards of radiomics.Keywords
This publication has 31 references indexed in Scilit:
- Quantitative Measurements Versus Receiver Operating Characteristics and Visual Grading Regression in CT Images Reconstructed with Iterative ReconstructionAcademic Radiology, 2018
- Influence of CT acquisition and reconstruction parameters on radiomic feature reproducibilityJournal of Medical Imaging, 2018
- Cancer imaging phenomics toolkit: quantitative imaging analytics for precision diagnostics and predictive modeling of clinical outcomeJournal of Medical Imaging, 2018
- Impact of image preprocessing on the volume dependence and prognostic potential of radiomics features in non-small cell lung cancerTranslational Cancer Research, 2016
- Quantitative Features of Liver Lesions, Lung Nodules, and Renal Stones at Multi–Detector Row CT Examinations: Dependency on Radiation Dose and Reconstruction AlgorithmRadiology, 2016
- Radiomics: Images Are More than Pictures, They Are DataRadiology, 2016
- Image quality in CT: From physical measurements to model observersPhysica Medica, 2015
- Measuring Computed Tomography Scanner Variability of Radiomics FeaturesInvestigative Radiology, 2015
- Comparing five different iterative reconstruction algorithms for computed tomography in an ROC studyEuropean Radiology, 2014
- Quantitative Statistical Methods for Image Quality AssessmentTheranostics, 2013