CT brain perfusion: A static phantom study of contrast‐to‐noise ratio and radiation dose
- 27 November 2016
- journal article
- research article
- Published by Wiley in Journal of Medical Imaging and Radiation Oncology
- Vol. 61 (3) , 361-366
Abstract
Introduction: Computed tomography perfusion (CTP) is increasingly employed in the diagnosis and management of ischaemic stroke but radiation dose can be significant and optimising contrast‐to‐noise ratio (CNR) is challenging. This study aimed to quantify and optimise the balance between CNR as a surrogate for image quality and radiation dose.Methods: A perspex head phantom with vials of dilute contrast agent was scanned using a Siemens Definition Flash 128‐slice scanner. The CTP protocol exposure parameters were adjusted over 70–120 kVp and 150–285 mAs. Measurements were obtained for the average dose per slice, Hounsfield Units (HU) for iodinated contrast agent, and the image noise for background regions of perspex. The CNR was measured as a function of the volumetric CT dose index (CTDIvol) and kVp.Results: A change from 120 to 80 kVp, achieved the same CNR with 60% reduction in dose. Alternatively, for the same dose, the change from 120 to 80 kVp improved CNR by +58%. A change from 80 to 70 kVp while operating at the same CNR, led to 13% reduction in dose. Alternatively, maintaining the same dose while changing from 80 to 70 kVp improved the CNR by +7%.Conclusion: Lower beam energies achieved the same CNR with less dose, or improved CNR at the same dose. A reduction from 80 kVp to 70 kVp may be clinically useful to optimise CTP acquisitions.Keywords
This publication has 14 references indexed in Scilit:
- Limited Reliability of Computed Tomographic Perfusion Acute Infarct Volume Measurements Compared With Diffusion-Weighted Imaging in Anterior Circulation StrokeStroke, 2015
- Radiation dose and image quality of 70 kVp cerebral CT angiography with optimized sinogram-affirmed iterative reconstruction: comparison with 120 kVp cerebral CT angiographyEuropean Radiology, 2015
- Cerebral computed tomography angiography using a 70 kVp protocol: improved vascular enhancement with a reduced volume of contrast medium and radiation doseEuropean Radiology, 2014
- Whole-Brain Adaptive 70-kVp Perfusion Imaging with Variable and Extended Sampling Improves Quality and Consistency While Reducing DoseAmerican Journal of Neuroradiology, 2014
- Improvement of image quality and radiation dose of CT perfusion of the brain by means of low-tube voltage (70 KV)European Radiology, 2014
- Comparison of Computed Tomography Perfusion and Magnetic Resonance Imaging Perfusion-Diffusion Mismatch in Ischemic StrokeStroke, 2012
- Perfusion computer tomography: imaging and clinical validation in acute ischaemic strokeBrain, 2011
- FDA Investigates the Safety of Brain Perfusion CTAmerican Journal of Neuroradiology, 2009
- Radiation Protection in MedicineAnnals of the ICRP, 2007
- CT-Expo - ein neuartiges Programm zur Dosisevaluierung in der CTRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2002