A Nonparametric Surrogate-Based Test of Significance for T-Wave Alternans Detection
- 19 April 2010
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 58 (5) , 1356-1364
Abstract
We present a nonparametric adaptive surrogate test that allows for the differentiation of statistically significant T-wave alternans (TWA) from alternating patterns that can be solely explained by the statistics of noise. The proposed test is based on estimating the distribution of noise-induced alternating patterns in a beat sequence from a set of surrogate data derived from repeated reshuffling of the original beat sequence. Thus, in assessing the significance of the observed alternating patterns in the data, no assumptions are made about the underlying noise distribution. In addition, since the distribution of noise-induced alternans magnitudes is calculated separately for each sequence of beats within the analysis window, the method is robust to data nonstationarities in both noise and TWA. The proposed surrogate method for rejecting noise was compared to the standard noise-rejection methods used with the spectral method (SM) and the modified moving average (MMA) techniques. Using a previously described realistic multilead model of TWA and real physiological noise, we demonstrate the proposed approach that reduces false TWA detections while maintaining a lower missed TWA detection, compared with all the other methods tested. A simple averaging-based TWA estimation algorithm was coupled with the surrogate significance testing and was evaluated on three public databases: the Normal Sinus Rhythm Database, the Chronic Heart Failure Database, and the Sudden Cardiac Death Database. Differences in TWA amplitudes between each database were evaluated at matched heart rate (HR) intervals from 40 to 120 beats per minute (BPM). Using the two-sample Kolmogorov-Smirnov test, we found that significant differences in TWA levels exist between each patient group at all decades of HRs. The most-marked difference was generally found at higher HRs, and the new technique resulted in a larger margin of separability between patient populations than when the SM or MMA were applied to the same data.Keywords
This publication has 25 references indexed in Scilit:
- An artificial vector model for generating abnormal electrocardiographic rhythmsPhysiological Measurement, 2010
- Effect of Noise on T‐Wave Alternans Measurement in Ambulatory ECGs Using Modified Moving Average versus Spectral MethodPacing and Clinical Electrophysiology, 2009
- The ABCD (Alternans Before Cardioverter Defibrillator) TrialJACC, 2009
- Adaptive Match Filter Based Method for Time vs. Amplitude Characterization of Microvolt ECG T-Wave AlternansAnnals of Biomedical Engineering, 2008
- Predicting Arrhythmia‐Free Survival Using Spectral and Modified‐Moving Average Analyses of T‐Wave AlternansPacing and Clinical Electrophysiology, 2007
- Characterization of Repolarization Alternans During Ischemia: Time-Course and Spatial AnalysisIEEE Transactions on Biomedical Engineering, 2006
- Methodological Principles of T Wave Alternans Analysis: A Unified FrameworkIEEE Transactions on Biomedical Engineering, 2005
- Microvolt T-Wave Alternans Distinguishes Between Patients Likely and Patients Not Likely to Benefit From Implanted Cardiac Defibrillator TherapyCirculation, 2004
- A dynamical model for generating synthetic electrocardiogram signalsIEEE Transactions on Biomedical Engineering, 2003
- Detecting periodicity in experimental data using linear modeling techniquesPhysical Review E, 1999