S-perindopril assay using a potentiometric, enantioselective membrane electrode

Authors:

Abstract
A potentiometric, enantioselective membrane electrode based on graphite paste (graphite powder and paraffin oil) has been constructed. The graphite paste is impregnated with a 10−3 mol/L 2-hydroxy-3-trimethylammoniopropyl-β-cyclodextrin (as chloride salt) solution. The potentiometric, enantioselective membrane electrode can be used reliably for enantiopurity tests of S-perindopril using a chronopotentiometric (zero current) technique, in the 10−5–10−2 mol/L concentration range (detection limit 5 × 10−6 mol/L), with an average recovery of 99.58% (RSD = 0.33%). The enantioselectivity was determined over R-perindopril and d-proline. The response characteristics of the enantioselective, potentiometric membrane electrode were also determined for R-perindopril. It was shown that l-proline is the main interfering compound. The surface of the electrode can be regenerated simply by polishing, obtaining a fresh surface ready to be used in a new assay. Chirality 11:631–634, 1999.