The Application of Integrated Force and Temperature Sensors to Enhance Orthotic Treatment Monitoring in Adolescent Idiopathic Scoliosis: A Pilot Study
Open Access
Abstract
Orthosis-wearing compliance is crucial for achieving positive treatment outcomes in patients with adolescent idiopathic scoliosis (AIS), for whom 23 h of daily wear is typically prescribed. However, self-reported compliance is subjective and often based on patients’ memory, leading to inaccuracies. While portable electronic devices have been developed to objectively monitor compliance, relying solely on temperature or force data can be insufficient. This study introduced a novel method that integrated both force and temperature data to estimate orthosis-wearing compliance. Twelve patients (eight females and four males) diagnosed with moderate AIS were included. Each patient was prescribed a thoracic-lumbar-sacral orthosis equipped with an integrated force and temperature sensor system. After one month of orthotic treatment, self-reported wear time averaged 17.8 ± 6.2 h/day, while the sensor indicated an average wear time of 13.3 ± 5.0 h/day. Most patients overestimated their compliance. Nighttime was the most common period for orthosis wear (6.1 h/day), whereas compliance during school hours (2.8 h/day) and after-school hours (3.7 h/day) was lower. The integration of force and temperature sensors provides a more comprehensive understanding of orthosis compliance. Future studies with larger samples and longer monitoring periods are needed to investigate the correlation between compliance and treatment outcomes.Keywords
This publication has 46 references indexed in Scilit:
- Prediction of future curve angle using prior radiographs in previously untreated idiopathic scoliosis: natural history from age 6 to after the end of growth (SOSORT 2022 award winner)European Spine Journal, 2023
- Effects of Mindfulness-Based Intervention to Improve Bracing Compliance in Adolescent Idiopathic Scoliosis Patients: a Randomized Controlled TrialMindfulness, 2023
- A smart point-of-care compliance monitoring solution for brace treatment of adolescent idiopathic scoliosis patientsSmart Health, 2021
- Compliance with night-time overcorrection bracing in adolescent idiopathic scoliosis: Result from a cohort follow-upMedical Engineering & Physics, 2020
- Clinically Significant Psychological and Emotional Distress in 32% of Adolescent Idiopathic Scoliosis Patients*Spine Deformity, 2018
- An advanced compliance monitor for patients undergoing brace treatment for idiopathic scoliosisMedical Engineering & Physics, 2015
- Measuring procedures to determine the Cobb angle in idiopathic scoliosis: a systematic reviewEuropean Spine Journal, 2013
- An objective measurement of brace usage for the treatment of adolescent idiopathic scoliosisMedical Engineering & Physics, 2010
- Electronic monitoring of scoliosis brace wear complianceJournal of Children's Orthopaedics, 2010
- Objective compliance of adolescent girls with idiopathic scoliosis in a dynamic SpineCor braceJournal of Children's Orthopaedics, 2010