CIGuide: in situ augmented reality laser guidance
Open Access
- 11 September 2019
- journal article
- research article
- Published by Springer Nature in International Journal of Computer Assisted Radiology and Surgery
- Vol. 15 (1) , 49-57
Abstract
Purpose : A robotic intraoperative laser guidance system with hybrid optic-magnetic tracking for skull base surgery is presented. It provides in situ augmented reality guidance for microscopic interventions at the lateral skull base with minimal mental and workload overhead on surgeons working without a monitor and dedicated pointing tools. Methods : Three components were developed: a registration tool (Rhinospider), a hybrid magneto-optic-tracked robotic feedback control scheme and a modified robotic end-effector. Rhinospider optimizes registration of patient and preoperative CT data by excluding user errors in fiducial localization with magnetic tracking. The hybrid controller uses an integrated microscope HD camera for robotic control with a guidance beam shining on a dual plate setup avoiding magnetic field distortions. A robotic needle insertion platform (iSYS Medizintechnik GmbH, Austria) was modified to position a laser beam with high precision in a surgical scene compatible to microscopic surgery. Results : System accuracy was evaluated quantitatively at various target positions on a phantom. The accuracy found is 1.2 mm ± 0.5 mm. Errors are primarily due to magnetic tracking. This application accuracy seems suitable for most surgical procedures in the lateral skull base. The system was evaluated quantitatively during a mastoidectomy of an anatomic head specimen and was judged useful by the surgeon. Conclusion : A hybrid robotic laser guidance system with direct visual feedback is proposed for navigated drilling and intraoperative structure localization. The system provides visual cues directly on/in the patient anatomy, reducing the standard limitations of AR visualizations like depth perception. The custom- built end-effector for the iSYS robot is transparent to using surgical microscopes and compatible with magnetic tracking. The cadaver experiment showed that guidance was accurate and that the end-effector is unobtrusive. This laser guidance has potential to aid the surgeon in finding the optimal mastoidectomy trajectory in more difficult interventions.Keywords
Funding Information
- Österreichische Forschungsförderungsgesellschaft (846056, 846056)
- Österreichische Forschungsförderungsgesellschaft (846056)
- Oesterreichische Nationalbank (16154)
This publication has 25 references indexed in Scilit:
- Stereoscopic depth perception in video see-through augmented reality within action spaceJournal of Electronic Imaging, 2014
- Evaluation of a portable image overlay projector for the visualisation of surgical navigation data: phantom studiesInternational Journal of Computer Assisted Radiology and Surgery, 2011
- OpenIGTLink: an open network protocol for image‐guided therapy environmentThe International Journal of Medical Robotics and Computer Assisted Surgery, 2009
- EPnP: An Accurate O(n) Solution to the PnP ProblemInternational Journal of Computer Vision, 2008
- The Image-Guided Surgery Toolkit IGSTK: An Open Source C++ Software ToolkitJournal of Digital Imaging, 2007
- Complete solution classification for the perspective-three-point problemIEEE Transactions on Pattern Analysis and Machine Intelligence, 2003
- Computer assistance for intraoperative navigation in ENT surgeryMinimally Invasive Therapy & Allied Technologies, 2003
- The distribution of target registration error in rigid-body point-based registrationIEEE Transactions on Medical Imaging, 2001
- Design and evaluation of a system for microscope-assisted guided interventions (MAGI)IEEE Transactions on Medical Imaging, 2000
- Closed-form solution of absolute orientation using unit quaternionsJournal of the Optical Society of America A, 1987