His bundle pacing guided by automated intrinsic morphology matching is feasible in patients with narrow QRS complexes
Open Access
- 4 March 2022
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 12 (1) , 1-12
Abstract
Pace mapping and visual comparison of the local pacing response with the intrinsic QRS morphology form the mainstay of His bundle pacing (HBP). We evaluated the performance of a surface lead morphology match algorithm for automated classification of the pacing response in patients with narrow intrinsic QRS undergoing electroanatomic mapping (EAM)-guided HBP. HBP was attempted in 43 patients. In 28 cases with narrow QRS, the EnSite AutoMap Module was used for automated assessment of the QRS morphology resulting from pace mapping in the His cloud area with either a diagnostic catheter or the His lead. An intrinsic morphology match score (IMS) was calculated for 1.546 QRS complexes and assessed regarding its accuracy and performance in classifying the individual pacing response as either selective HBP (S-HBP), nonselective HBP (NS-HBP) or right ventricular stimulation. Automated morphology comparison of 354 intrinsic beats with the individual reference determined a test accuracy of 99% (95% CI 98.96–99.04) and a precision of 97.99–99.5%. For His-lead stimulation, an IMS ≥ 89% identified S-HBP with a sensitivity, specificity and positive predictive value of 1.00 (0.99, 1.00) and a negative predictive value of 0.99 (0.98, 1.00). An IMS between 78 and < 89% indicated NS-HBP with a sensitivity and specificity of 1.00 (0.99, 1.00) and 0.99 (0.98, 1.00), respectively. IMS represents a new automated measure for standardized individual morphology classification in patients with normal QRS undergoing EAM-guided HBP.Clinical trial registration: NCT04416958.Keywords
This publication has 13 references indexed in Scilit:
- 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delayHeart Rhythm, 2019
- Direct Visualization of the His Bundle Pacing Lead Placement by 3-Dimensional Electroanatomic MappingCirculation: Arrhythmia and Electrophysiology, 2019
- Programmed His Bundle PacingCirculation: Arrhythmia and Electrophysiology, 2019
- Novel Method for Assessment of His Bundle Pacing Morphology Using Near Field and Far Field Device ElectrogramsCirculation: Arrhythmia and Electrophysiology, 2019
- Low Fluoroscopy Permanent His Bundle Pacing Using Electroanatomic MappingCirculation: Arrhythmia and Electrophysiology, 2019
- Permanent His bundle pacing: Recommendations from a Multicenter His Bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-upHeart Rhythm, 2018
- Electrophysiological observations of acute His bundle injury during permanent His bundle pacingJournal of Electrocardiology, 2016
- How to perform permanent His bundle pacing in routine clinical practiceHeart Rhythm, 2016
- Optimum threshold estimation based on cost function in a multistate diagnostic settingStatistics in Medicine, 2011
- Permanent His-bundle pacing: seeking physiological ventricular pacingEP Europace, 2010