In‐field and out‐of‐field microdosimetric characterisation of a 62 MeV proton beam at CATANA
Authors:
- 28 April 2021
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 48 (8) , 4532-4541
Abstract
Purpose: A 5 and 10 μm thin silicon on insulator (SOI) 3D mushroom microdosimeter was used to characterize both the in‐field and out‐of‐field of a 62 MeV proton beam.Methods: The SOI mushroom microdosimeter consisted of an array of cylindrical sensitive volumes (SVs), developed by the Centre for Medical Radiation Physics, University of Wollongong, was irradiated with 62 MeV protons at the CATANA (Centro di AdroTerapia Applicazioni Nucleari Avanzate) facility in Catania, Italy, a facility dedicated to the radiation treatment of ocular melanomas. Dose mean lineal energy, (), values were obtained at various depths in PMMA along a pristine and spread out Bragg peak (SOBP). The measured microdosimetric spectra at each position were then used as inputs into the modified Microdosimetric Kinetic Model (MKM) to derive the RBE for absorbed dose in a middle of the SOBP 2Gy (RBED). Microdosimetric spectra were obtained with both the 5 and 10 μm 3D SOI microdosimeters, with a focus on the distal part of the BP. The in‐field and out‐of‐field measurement configurations along the Bragg curve were modeled in Geant4 for comparison with experimental results. Lateral out‐of‐field measurements were performed to study secondary particles’ contribution to normal tissue’s dose, up to 12 mm from the edge of the beam field, and quality factor and dose equivalent results were obtained.Results: Comparison between experimental and simulation results showed good agreement between one another for both the pristine and SOBP beams in terms of and RBED. Though a small discrepancy between experiment and simulation was seen at the entrance of the Bragg curve, where experimental results were slightly lower than Geant4. The dose equivalent value measured 12 mm from the edge of the target volume was 1.27 ± 0.15 mSv/Gy with a value of 2.52 ± 0.30, both of which agree within uncertainty with Geant4 simulation.Conclusions: These results demonstrate that SOI microdosimeters are an effective tool to predict RBED in‐field as well as dose equivalent monitoring out‐of‐field to provide insight to probability of second cancer generation.Keywords
This publication has 24 references indexed in Scilit:
- The physics of proton therapyPhysics in Medicine & Biology, 2015
- Relative biological effectiveness (RBE) values for proton beam therapy. Variations as a function of biological endpoint, dose, and linear energy transferPhysics in Medicine & Biology, 2014
- Secondary neutron dose measurement for proton eye treatment using an eye snout with a borated neutron absorberRadiation Oncology, 2013
- Charged Particle Radiation Therapy for Uveal Melanoma: A Systematic Review and Meta-AnalysisInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2013
- Microdosimetric calculation of relative biological effectiveness for design of therapeutic proton beamsJournal of Radiation Research, 2012
- CATANA protontherapy facility: The state of art of clinical and dosimetric experienceThe European Physical Journal Plus, 2011
- Assessment of out‐of‐field absorbed dose and equivalent dose in proton fieldsMedical Physics, 2009
- Out-of-Field Dose Equivalents Delivered by Passively Scattered Therapeutic Proton Beams for Clinically Relevant Field ConfigurationsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Out‐of‐field dose equivalents delivered by proton therapy of prostate cancerMedical Physics, 2007
- Quantities used in radiological protectionAnnals of the ICRP, 1991