Technologies for Prolonging Cardiac Implantable Electronic Device Longevity
Author:
- 15 January 2017
- journal article
- review article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 40 (1) , 75-96
Abstract
Prolonged longevity of cardiac implantable electronic devices (CIEDs) is needed not only as a passive response to match the prolonging life expectancy of patient recipients, but will also actively prolong their life expectancy by avoiding/deferring the risks (and costs) associated with device replacement. CIEDs are still exclusively powered by nonrechargeable primary batteries, and energy exhaustion is the dominant and an inevitable cause of device replacement. The longevity of a CIED is thus determined by the attrition rate of its finite energy reserve. The energy available from a battery depends on its capacity (total amount of electric charge), chemistry (anode, cathode, and electrolyte), and internal architecture (stacked plate, folded plate, and spiral wound). The energy uses of a CIED vary and include a background current for running electronic circuitry, periodic radiofrequency telemetry, high‐voltage capacitor reformation, constant ventricular pacing, and sporadic shocks for the cardiac resynchronization therapy defibrillators. The energy use by a CIED is primarily determined by the patient recipient's clinical needs, but the energy stored in the device battery is entirely under the manufacturer's control. A larger battery capacity generally results in a longer‐lasting device, but improved battery chemistry and architecture may allow more space‐efficient designs. Armed with the necessary technical knowledge, healthcare professionals and purchasers will be empowered to make judicious selection on device models and maximize the utilization of all their energy‐saving features, to prolong device longevity for the benefits of their patients and healthcare systems.Keywords
Funding Information
- Boston Scientific Corporation
- St. Jude Medical
- Brookhaven National Laboratory
This publication has 69 references indexed in Scilit:
- Mortality of Patients with Heart Failure and Reduced Ejection Fraction (HFrEF) Who Receive Either ICD or CRT-D Has Improved Yearly from 2003 to 2010: The Altitude RegistryJournal of Cardiac Failure, 2013
- Adsorption of Single Li and the Formation of Small Li Clusters on Graphene for the Anode of Lithium-Ion BatteriesACS Applied Materials & Interfaces, 2013
- Rate, causes, and impact on patient outcome of implantable device complications requiring surgical revision: large population survey from two centres in ItalyEP Europace, 2013
- A Wireless Magnetic Resonance Energy Transfer System for Micro Implantable Medical SensorsSensors, 2012
- Effect of Early Diagnosis and Treatment With Percutaneous Lead Extraction on Survival in Patients With Cardiac Device InfectionsThe American Journal of Cardiology, 2012
- Timing of the Most Recent Device Procedure Influences the Clinical Outcome of Lead-Associated EndocarditisJACC, 2012
- Acute Clinical Evaluation of a Left Ventricular Automatic Threshold Determination Algorithm Based on Evoked Response SensingPacing and Clinical Electrophysiology, 2011
- Impact of ICD Battery Longevity on Need for Device Replacements-Insights from a Veterans Affairs DatabasePacing and Clinical Electrophysiology, 2010
- Update on Cardiovascular Implantable Electronic Device Infections and Their ManagementCirculation, 2010
- A Cardiac Evoked Response Algorithm Providing Threshold Tracking: A North American Multicenter StudyPacing and Clinical Electrophysiology, 2000